0000137242 Rev.5.1
10
At the end of each calibration point, the instrument will display the quality of the electrode as a percentage.
3.3.2 ORP (REDOX) PROBE CALIBRATION
Connect the ORP probe to the instrument as indicated in the electrical connections.
Select the ORP probe from the Calibration menu.
Select whether to perform the calibration in Automatic (
AUTO
) or Manual (
MAN
) mode.
At the end of each calibration point, the instrument will display the quality of the electrode as a percentage.
pH CAL. Type: Man
pH CAL. Type: Man
7.01pH 25.0°C
pH CAL. Type: Man
7.00pH 25.0°C
Wait 60”
pH CAL. Type: Man
4.01pH 25.0°C
Wait 60”
ORP CAL. Type: Auto
ORP CAL. Type: Auto
+475mV
ORP CAL. Type: Auto
+475mV
Wait 60”
MAN
In Manual (
MAN
) mode:
•
Immerse the probe in the first solution, insert
the solution’s pH value and press
Enter
•
Wait 60 seconds. When finished, the
instrument will display the probe’s quality as
a percentage value.
•
Immerse the probe in the second solution
and insert the solution’s pH value..
•
Wait 60 seconds. When finished, the
instrument will display the probe’s quality as
a percentage value.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.
AUTO
In Automatic (
AUTO
) mode:
•
Immerse the probe in the +475mV solution
and press
Enter
•
Wait 60 seconds. When finished, the
instrument will display the probe’s quality as
a percentage value.
•
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.