0000137242 Rev.5.1
26
4.6 Protocollo ModBus RTU
Di seguito la lista comandi per la trasmissione dati via Porta Seriale RS485 RTU
> Read Holding Register
> Write Multiple Register
> Report Slave ID
In risposta al comando Report Slave ID il sistema risponde con una stringa:
"Device"+ Firmware code + Edition
ModBus profile table Address
Desctiption
Property
Range
Operating state
1000
pH Measure
Read 0 to 1400
Full Operating
1001
Cl Measure
Read 0 to 500
Full Operating
1002
Orp Measure
Read -2000 to +2000
Full Operating
1003
Temperature Measure
Read 0 to 1050
Full Operating
1004
Frequency Measure
Read 5 to 15000
Readable but not active
1005
Conductivity Measure
Read 0 to 10000
Readable but not active
1006
Pot1 Measure
Read 0 to 20000
Readable but not active
1007
Pot2 Measure
Read 0 to 20000
Readable but not active
1008
Status
Read See Note 1
Full Operating
1100
SetPoint pH
Read/Write 0 to 1400
Full Operating
1101
SetPoint Cl
Read/Write 0 to 500
Full Operating
1102
SetPoint Orp
Read/Write -2000 to +2000
Full Operating
1103
SetPoint Temperature
Read/Write 0 to 1050
Full Operating
Note1: STATUS (bit field 16 bit register)
1
LEVEL_0_ALARM
2
LEVEL_1_ALARM
3
HOLD_ALARM
4
REED_ALARM
5
Reserverd
6
Reserverd
7
Reserverd
8
Reserverd
9
Reserverd
10 Reserverd
11 Reserverd
12 Reserverd
13 STAT (Wieving statistics Mode)
14 Reserverd
15 OFA (Over Feed Alarm)
16 PERMANECY (Permanency Allarm)