0000137242 Rev.5.1
18
•
Menu index “3A2” Frequency output proportional to pH measurement (FWM PH)
The frequency output (Open collector circuit) can be used to control and guide the dosing of a remote
system in proportion to the pH measurement.
•
Menu index “3A3” Current output proportional to pH measurement (OUT mA PH)
Note:
The value set under the
HOLD mA
item is automatically generated by the instrument when a functional Hold
is present, for example due to a lack of water Flow Alarm or an enabled Voltage Input.
•
Menu index “3A4” PH ALARMS
Note:
The
Holding Range
and
Holding Time
items must be used together.
The indicated function controls the chemical measurement at a constant value for long periods of time. This alarm
can help to prevent incorrect dosing as a result of damaged probes.
Item
Default Value
Range
Standard FWM:
SetPoint:
7.20 pH
0-14 pH
Dose Type:
Acid
Acid / Alka
Pulse:
20 pulses/minute
20-150 pulses/minute
Proportional Band:
0.3 pH
0.3-3pH
Item
Default Value
Range
Standard Out mA:
Range 0/4-20mA:
4-20 mA
0-20 mA or 4-20 mA
Start (4mA):
0 pH
0.00 - 14.00 pH
End (20mA):
14 pH
14.00 - 0.00 pH
Hold mA:
4 mA
0-20 mA
Item
Default Value
Range
pH Alarms List
Alarm minimum:
6.2 pH
0-14 pH
Alarm Maximum:
8.2 pH
0-14 pH
OFA (Maximum dose timer)
Off
10-3600 Seconds
Holding range:
Off
0.2-3 pH
Holding time:
Off
10-3600 Seconds
Level Alarm: System stop or
alarm display
Disabled
Enabled/Disabled
3A2 FWM PH
SET POINT: 7.20pH
DOSE TYPE: ACID
PULSE: 20/min
↓
3A2 FWM PH
DOSE TYPE: ACID
↑
PULSE: 20/min
PROP. BAND: 0.30pH
3A3 OUT mA PH
RANGE: 4-20 mA
START( 4): 0.00pH
END (20): 14.00pH
↓
3A3 OUT mA PH
START( 4): 0.00pH
↑
END (20): 14.00pH
HOLD mA: 4.00mA
3A4 PH ALARMS
MIN VAL.: 6.20pH
MAX VAL.: 8.20pH
OFA: OFF
↓
3A4 PH ALARMS
HOLDING RANGE: OFF
↑
HOLDING TIME: OFF
LEVEL ALARM: DISABLED