MANUAL DE INSTRUCCIONES CODIGO 80303 REV B Sep2019 (Sujetas a modificaciones sin previo aviso) Pag.: 19
J
.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
8.5 Alarms
When entering this setting, screen Fig 8.5 appears. Use the
Up/Down
arrows
to select the value of over temperature or low temperature you want for the
temperature range.
This setting is simply a warning, it does not stop the operating cycle. This notice
will be reflected in the memory log.
Press the
BACK
key to return to the settings menu or the
HOME
key to return
to the initial screen.
8.6 Temperature Offset Adjustment
The Offset adjustment is protected by a password to prevent any user from
modifying this parameter.
Press the
CAL
button to access this setting.
If the
CAL
button is pressed without entering the password, it will indicate that
a valid password value is necessary to continue with the action. Therefore, it is
necessary to select
ENTER PASSWORD
option before pressing the
CAL
button.
The default password value for the calibrator is “
0000
”.
Press on the temperature Offset parameter and use the
Up/Down
keys to
modify this value.
Press the
BACK
key to return to the settings menu or the
HOME
key to return
to the initial screen.
The password value remains valid until the oven is switched off and on again.
Perform this action to prevent another user from modifying the value.
8.7 Password Change
If you want to change the factory password, use this menu and follow the ins
-
tructions that appear to change the password.
Press the
BACK
key to return to the settings menu or the
HOME
key to return
to the initial screen.
8.8 Logs Download
To download the data stored in the control, select the option
LOGS DOWNLOAD
.
Screen Fig. 8.7 will appear.
Connect a memory pen drive to the front connector (4) and follow the instructions
on the screen.
There are two download options: either download all logs or the last 10 only.
During the loading of data in the USB memory, information about % of loading
appears.
Press the
BACK
key to return to the settings menu or the
HOME
key to return
to the initial screen.
Fig. 8.6 Password entry
Fig. 8.5 Setting alarm margins
Fig. 8.6.1 Temperature Offser modification
Fig. 8.7 Password modification
Fig. 8.8 Logs download