34
_SYSTEM SETUP
Return to SYSTEM SETUP screen by pressing NEXT button in key pad.
PRINT POSITION
It adjusts the printing position in the direction of U-D (up-down) and L-R (left-right) to fit to the pre-
formatted result sheet.
Select
(PRINT POSITION) on SYSTEM SETUP screen with ‘◀’ and ‘▶’ button and press SET
button in key pad.
Pre-set: 00 for U-D and 00 for L-R
Range: 99 for U-D and 99 for L-R
Choose
“
BCA
”
by pressing
‘
1
’
in key pad. Choose
“
CHD
”
by pressing
‘
2
’
in key pad.
Choose U-D by pressing
‘
3
’
in key pad. Choose L-R by pressing
‘
4
’
in key pad.
Pressing ▶ button moves print position down or right.
Pressing ◀ button moves print position up or left.
Every single press moves print position by about 0.2 mm.
Press BACK button in key pad to save it.
Return to SYSTEM SETUP screen by pressing NEXT button in key pad.
Note
1.
L-R (left-right): - is moving to the left and + is moving to the right.
U-D (up-down): - is moving up and + is moving down.
2.
The print position needs to be separately set on the result sheet for Body composition
analysis and result sheet for children each.
Press
“
BCA
”
and adjust the print position of Body composition analysis result sheet
and then press
’
BACK
’
button to save the setting.
Press
“
CHD
”
and adjust the print position of result sheet for children and then press
‘
BACK
’
button to save the setting.