SELVAS Healthcare, Inc.
62
_ERROR & REPAIR
ERROR & REPAIR
Kinds of Error & Repair
Error
Cause
Repair
Out of range of
impedance
When the subject
’
s body impedance
is deviated from the limit
Insufficient touch to electrodes
Impedance is out of range
Range: 100 ~ 950
Ω
Clean the measuring parts (the
electrodes, palms, and soles) and try
again.
Measure again with correct posture.
Do not move during measurement.
If the same error is repeated, please
contact our company or its local
distributor where this device is
purchased.
Out of range of
body fat
When the subject
’
s P.B.F. is deviated
from the limit
Incorrect input of personal data
P.B.F. is out of range
Clean the Handle electrodes and try
again.
If the same error is repeated, please
contact our company or its local
distributor where this device is
purchased.
Out of range of
measurem-ent
When the subject
’
s fatness is
deviated from the limit
Mechanical error
Input height correctly.
Check the weight is correct and try
again..
If the same error is repeated, please
contact our company or its local
distributor where this device is
purchased.
Can
’
t input the
height
When the subject
’
s height is
deviated from the limit
Incorrect input of height
Input height correctly. If the
subject
’
s height is out of the range,
height can't be entered.
Can
’
t measure the
weight
When the subject
’
s weight is
deviated from the limit
Measuring error
Measure weight again. If the
subject's weight is out of range,
weight can't be measured.
Summary of Contents for ACCUNIQ BC300
Page 1: ...English 06 2021 V3 01 User Manual BC300...