SELVAS Healthcare
54
_ERROR & REPAIR
ERROR & REPAIR
Kinds of Error & Repair
Error
Cause
Repair
Out of range
of impedance
When the subject’s body
impedance deviates from the
limit
Insufficient touch to
electrodes
Impedance is out of
range
Range: 100 ~ 950 Ω
Clean the measuring parts (the electrodes,
palms, and soles) and try again
Measure again with correct posture
Do not move during measurement
If the same error is repeated, please contact
SELVAS Healthcare or its local distributor from
where this device is purchased
Out of range
of body fat
When the subject’s P.B.F.
deviates from the limit
Incorrect input of
personal data
P.B.F. is out of range
Clean the electrode holders and try again.
After checking that there is neither something
with wrong input of personal data (age,
gender) nor with measuring error of weight
and impedance, try again
It can't measure if the P.B.F. is out of range
When the same error occurs even after re-
measurement, please contact SELVAS
Healthcare or its local distributor from where
the device was purchased
Out of range
of
measurement
When the subject’s fatness is
deviated from the limit
Mechanical error
Input height correctly or if installed height
already, measure again
Confirm to measure weight and try again
correctly
It can't measure if the fatness is out of range.
When the same error is occurred even re-
measurement, please contact with SELVAS
Healthcare or its local distributor where is
purchased
Can’t input the
height
When the subject’s height is
deviated from the limit
- Incorrect input of height
Input height correctly. If the subject’s height is
out of range, height can't be entered