11)
Dry hands and feet might affect the results. In case your hands and feet are dry, please wipe them
with wet wipes before measuring for enhanced electrical conductivity.
In order to keep one's health and the balance of body composition, check the changes of body composition
through the continuous analysis and compare the results. Make sure that the body composition should be
measured under the same physical and environmental conditions. If the condition before the measurement
such as volume of a meal, meal time, and activities (exercise, sauna, drinking lots of beverage, urination,
etc.) are kept same, the reproducibility of a device is obtained. Therefore, the data can be used to evaluate
the change of body composition.
Correct Posture
How to Touch Plate Electrodes
Make sure that the plate electrodes are clean.
Take off the socks or stockings then, stand on the plate electrodes.
Remove sweat or foreign matters on the soles.
Fairly place the bare feet on the plate electrodes. Make sure that the clothes are not between the
soles and the plate electrodes.
(O) Correct stand
(X) Imbalanced touch
(X) Foreign matters
How to Touch Handle Electrodes
Grip handle electrodes with fingers and palms.
4 electrodes should be touched impartially.
Stretch both arms and spread them 30
°
from the body.