SELVAS Healthcare
40
_MEASUREMENT AND ANALYSIS
(O) Correct grip
(X) Imbalanced touch
(X) Grip with only palm
Note
1.
When a subject has small hands or feet and cannot cover all electrodes sufficiently,
please pay attention to touch all electrodes fairly. The way to touch the electrodes will
affect the reliability of the results.
2.
During the measurement, the subject should not be touched by others or conductive
materials.
3.
If all eight electrodes are not perfectly touched during measurement, measurement
will be stopped or the result is not reliable.
Measuring Posture
Step the scale in the bare feet. Stretch both arms and spread them 30
°
from the body.
Press start buttons with thumbs for 2 ~ 3 seconds to start the measurement. Once it starts, release
the start button and hold the same posture until the measurement is over.
Do not speak or move the body until the measurement is completed.
Do not bend or shake the arms until the measurement is completed.
The measurement will be stopped if all eight electrodes are not fairly touched.
<Measuring posture for Whole and Upper body>
<Measuring posture for lower body>