16
Magnetic field
of commercial
frequency
(50/60Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
hospital
environment.
Note
U
T is the a.c. mains voltage prior to application of the test level.
3) Guidance and manufacturer’s declaration – electromagnetic immunity 2
The ACCUNIQ BC380 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ACCUNIQ BC380 should assure that it is used in such an environment.
Immunity test
IEC
60601
test
level
Compliance
level
Electromagnetic
environment-
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
3 Vrms
3 V/m
Portable
and
mobile
RF
communications equipment should
be used no closer to any part of the
ACCUNIQ
BC380,
including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended
separation
distance
d =
1.2
d =
1.2
80 MHz to 900 MHz
d =2.3
900 MHz to 2,5 GHz
where
P
is the maximum output
power rating of the transmitter in
watts
(W)
according
to
the
transmitter manufacturer and
d
is
Summary of Contents for Accuniq BC380