18
4) Recommended separation distances between portable and mobile RF communications
equipment and the ACCUNIQ BC380
The ACCUNIQ BC380 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ACCUNIQ BC380 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the ACCUNIQ BC380 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =
1.2
80 MHz to 900 MHz
d =
1.2
900 MHz to 2,5 GHz
d =
1.2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note
1. At 80 MHz and 900 MHz, the separation distance for the higher frequency
range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.