55
STORAGE & MAINTENANCE
1) Pay attention to the allowable value of the electric current.
2) Avoid direct sunlight, humidity, dust, thick oil, salty air or extreme changes in temperature.
3) Do not install or store the device in a place where chemicals or gas are stored.
4) Do not use the device in unstable environments with a high amount of vibrations or
heavy impacts.
5) Connect the ground located on the backside of this device to the terminal plate to prevent any
electric shock from power surges or other electrical current changes.
6) Do not place heavy objects on or drop anything on to the device, and avoid strong
impacts.
7) Do not disassemble or modify the device.
8) If the unit has not been used for an extended period, confirm with an expert that all functions and
physical mechanisms are in good condition before use.
9) Do not introduce any liquid on to the device or insert any foreign substances.
10) If foreign substances are introduced, or if the device is exposed to harmful environments, the
unit must be examined by a qualified technician before use.
11) Use only the power cable, adapter, and fuses provided by SELVAS Healthcare.
Please confirm the covering of the cable, the state of the adapter connection, and other safety
checks as below:
• RS 232C cable • USB port • Adapter
12) When disconnecting the power cable, turn off the power switch first then unplug the unit.
13) Store the unit in an environment with an ambient Temperature -25 ~ 70 °C, Humidity lower
than 93 % (non condensing)
14) The operating environment should have an ambient Temperature 5 ~ 40 °C, Humidity 15 ~
93 % (non condensing)
15) Do not store or use this device in environments under 70 kPa (700 mbar) or over 106
kPa (1060 mbar) of atmospheric pressure.
16) Cleaning & Disinfection
①
Cleaning: Use a soft gauze cloth with volatile liquid like alcohol (Ethyl or Isopropyl alcohol
70~90%) to clean the unit.
Clean the device every 2~3days. Do not use a wet cloth.
②
Disinfection: After measurement, use a soft gauze cloth with volatile liquid like benzene and
alcohol. Then, wipe the enclosure with a soft fiber cloth. Please wipe after every measurement for
electrode disinfection.
17) Please refer to and abide by the “SAFETY PRECAUTIONS.”
Caution
Users must be sure to use sterile safety equipment such as gloves when in contact with
or cleaning electrodes.
SELVAS Healthcare is not responsible for safety accidents caused by users’
carelessness.