56
ERROR & REPAIR
1. Troubleshooting& Repair
Error
Cause
Repair
Out of range
of
impedance
When the subject’s body impedance
deviates from the limit
- Insufficient contact with electrodes
- Impedance is out of range
-
Range: 100 ~ 950 Ω
• Clean the measurement contact areas (the
electrodes, your palms and soles) and try
again.
• Measure again with correct posture.
• Do not move during measurement.
• If the same error is repeated, please contact
our company or the local distributor from which
this device was purchased.
Out of range
of body fat
When the subject’s P.B.F. deviates from
the limit
- Incorrect input of personal data
- P.B.F. is out of range
• Clean the electrode holders and try again.
• After checking that the personal data is correct
(age, gender) and that there are no measuring
errors with weight and impedance, try again.
• Measurement cannot proceed if the P.B.F. is
out of range.
• When the same error occurs even after re-
measurement, please contact our company or
the local distributor from which the device was
purchased.
Out of range
of
measureme
nt
When the subject’s fatness is deviated
from the limit
- Mechanical error
• Input height correctly or if installed height
meter, measure again.
• Confirm weight measurement and try again.
• Measurement cannot proceed if the fatness is
out of range.
• When the same error occurs even after re-
measurement, please contact our company or
the local distributor from which the device was
purchased.
Insufficient
electrode
connection
to Ankle
• When even one of the 4 electrodes does
not sufficiently connect.
• Reconnect all electrodes again.
Can’t input
the height
When the subject’s height is deviated
from the limit
- Incorrect input of height
• Input height correctly. If the subject’s height is
out of range, height can't be entered.
Can’t
measure the
weight
When the subject’s weight deviates from
the limit
- Measuring error
- Moving during the measurement
• Measure the weight again. Don't move or
speak during measurement.
• The weight can't be measured if it is out of
range.
• When the same error occurs even after re-
measurement, please contact our company or
the local distributor from which the device was
purchased.