57
2. Error & Repair
Error code
Cause
Repair
38001
• Electrode and
measuring parts are not
detected or dirty
• Faulty in impedance
measuring
• Try to measure again after cleaning the
electrode holders with soft gauze.
• Try again after cleaning the hands and soles.
• Check if foreign materials are between
electrodes and measuring parts.
• Try again in correct posture and hold the
electrodes according to the measuring method.
38002
• Ankle electrode and
measuring parts are not
detected
• Try to measure again after cleaning the ankle
electrode with soft gauze.
• Try again after cleaning the ankle electrode.
38003
• There is an error in
measured impedance by
external influence.
• Check installation condition (AC cord or around
equipment
• Contact our company or distributor.
38004
• Out of measurement
range of impedance
• range: 100~950Ω
• Try again after cleaning the hands and soles.
• Check if foreign materials are between
electrodes and measuring parts.
• It can’t measure if the impedance is out of
range.
38005
• Out of measurement
range of PBF
• range: Under 75%
• Try again after cleaning the hands and soles.
• It can’t measure if the PBF is out of range.
38006
• Continuous error for 3
times
• Step off and try again.
• Contact our company or distributor.
38007
• Disconnect Color
board and Main board
• Contact our company or distributor.
Summary of Contents for Accuniq BC380