24
ENAMEL RATER MANUAL: V.10
4.9
AUTOSTART
When SW1/3 on the micro board (see Appendix A Fig 1) is set off
the operation of the changes as follows:
1.
The semi-automatic operation described in Para 1.2.2.
does not operate. After placing a can in the can stand the
operator must press start and ensure that the level and
contact check indicators are off before enamel rating will
commence.
Note that if either SW11/5 or SW11/6 is off then the
autostart function is disabled, and rating will commence
only when START is pressed.
4.10 AUTOSAVE
When SW1/4 on the micro board (see Appendix A Fig 1) is set on
the operation of the changes as follows:
1.
The operates in a "fully automatic" mode. After
each can has been rated the displayed reading is
automatically saved as the can is removed from the can
stand and the contact check and level check indicators
are illuminated. The combination of autosave and
autostart results in a "fully automatic" operation.