The Menu
31
Please note, that the alarm is always referred to the absolute water level.
Sensor Type
Select here the type of water level sensor you have connected. The water level probe PS has
a range of 3.0 m (Type: PS-300-MA) while the EPS exhibits a range of 2.5 m (Type: EPS-250-
MA or EPS-250-MAK).
Plug name
With plug name you can give the sensor a name of max. 8 characters, which will be displayed
in the SENECT Control App.
Change zero-point
If the sensor is used for refilling, you can change the zero-point (reference level) so that it is
shiftet by the manually set level.
S2: O2S Name (Example Oxygen Sensor O2S)
Sensor calibration
Every sensor should be calibrated regularly to achieve best accuracy. In this menu item, you
can calibrate either the oxygen sensor signal or the temperature signal of the O2S.
Please choose, which parameter should be calibrated: e.g. O2 % a.s. for calibrating the oxygen
signal.
All calibrations are programmed as 2-point calibrations, where also only one point can be
changed. This is for example important if you want to calibrate your oxygen sensor only in air.
To do so, please hold the sensor
in air or 100% oxygen saturated
water and select the upper
calibration point 2 (100% a.s.). If
the measurement value is stable,
press ok.
For the 2-point calibration, you
have also to change calibration
point 1 (0%-oxygen). Insert the
sensor in 0%-oxygen solution
and wait until the value is stable.
Confirm the calibration point
with ok.
Fig. 10: 2-point calibration of the oxygen sensor.