background image

06 

  At Gas Measure Mode, Peak value is displayed (Peak values of all 4 sensors) by 

pressing  Arrow  key  for  3  seconds,  the  value  of  TWA  (only  for  CO  and  H2S)  is 

shown  by  pressing  it  again,  and  that  of  STEL  (only  for  CO  and  H2S)  is  indicated 

consecutively by pressing it once more. The program returns to Gas Measure Mode 

by pressing Power key at the state indicating Peak, TWA and STEL.  And it reverts 

automatically  to  Gas  Measure  Mode  as  well  without  using  the  button  for  30 

seconds at the state of Peak, TWA and STEL. 

  The  instrument  is  converted  to  Gas  Measure  Mode  as  below, 

when  power  is  turned  on.  Gas  Measure  Mode  indicates  the 

concentration  of  gas  and  the  remained  capacity  of  battery  on 

LCD,  and  it  displays  the  gas  concentration  of  Oxygen  by  %, 

Inflammable Gas by %LEL, H2S and CO Gas by ppm units. In the 

event that the concentration of the specific gas changes, it indica 

-tes the value of concentration in real-time. If it exceeds A1 or A2 Alarm standard 

(or STEL / TWA), the measured value and A1 or A2 (icon in case of STEL / TWA) is 

displayed at one second interval in turn alerting by alarm sound and vibration. 

  As of A1, the instrument program will be converted to the original condition and 

starts the operation, when the concentration value of the measured gas returns to 

the normal state. (LED and LCD Backlight are displayed continuously and the alarm 

sound  stops,  if  power  key  is  pressed  during  the  alarm  sounds.  During  the  alarm 

sound, the measured value and A1 or A2 are displayed in turn on LCD.)  In case of 

A2,  power  should  be  disconnected  in  order  to  stop  the  alarm.  When  the 

concentration value of the measured gas exceeds the maximum measuring range, it 

is indicated as the maximum value.  And LED, alarm sound and vibration applicable 

to A2 Alarm Standard operate together 

4. Operation Method 

Detection Mode 

Indication of Peak Value and TWA & STEL Value 

Password Mode 

  Password  input  screen  appears  by  pressing  Power  key  for  5  seconds.  Number 

changes from far left to the right side. It moves upward by each 1 step and blinks 

by  pressing  Arrow  key,  and  the  blinking  2nd  number  moves  the  position  by 

pressing Power key. If key. 

Summary of Contents for SP12C7

Page 1: ...MODEL SP12C7 Portable 4 gas detector Operating Manual...

Page 2: ...Guarantee and Repair Senko Co Ltd guarantees the products of SP series for 24 months from the shipping date and repairs or replaces the defected product during warranty period at no cost Nevertheless...

Page 3: ...mes of Exterior and Display 04 Start and End of Operation 05 Operation Method 06 Backlight 07 Start and End of Sensor 08 Calibration 08 Method of Alarm Set and Display 10 Battery Sensor replacement 13...

Page 4: ...y Digital LCD Audible Alarm 90dB Visual Alarm Red LED Vibrator Alarm Vibration Alarm Power Alkaline battery 2EA Rechargeable NI MH battery pack Applicable Temperature Humidity 20 50 10 95 RH non conde...

Page 5: ...sensor 2 Buzzer 3 LCD display 4 On Off Key 5 Key 6 Alarm LED 7 Fastening clip 8 Type label 9 Charging contacts 10 Battery pack 1 2 3 4 5 6 7 8 9 10 04 LCD display symbols 2 Names and Functions of Ext...

Page 6: ...en the instrument is stabilized after the display of the above set values it converts to Gas Detection Mode displaying the mark At normal status the mark disappears after 30 seconds and the concentrat...

Page 7: ...WA the measured value and A1 or A2 icon in case of STEL TWA is displayed at one second interval in turn alerting by alarm sound and vibration As of A1 the instrument program will be converted to the o...

Page 8: ...word is put repeatedly at RE the password will be changed and the program will return Password Change Mode where icon is blinking If further change of password is not required return to Gas Measure Mo...

Page 9: ...s unless the button works 5 LCD Backlight Individual Sensor On Off 6 Start and End of Sensor cLr Mode is factory reset mode Factory reset mode prohibits user except factory from using of cLr Mode Plea...

Page 10: ...stored in the instrument Calibration is in general to be performed monthly or quarterly and can be adjusted according to frequency of the use In the event that calibration fails it would not convert...

Page 11: ...rators Save value Change value Start calibration If Standard Gas Calibration starts screen of detector is displayed sequentially according to gas kind That is calibration proceeds for 60 seconds as in...

Page 12: ...k with alarm sound In this case calibration of the sensor or aftercare of the instrument is required Gas Combustible Oxygen Carbon monoxide Hydrogen Sulfide Concentration 50 LEL CH4 18 vol 100 ppm 25...

Page 13: ...de when the change of Alarm Value of each gas is completed Alarm Set of Each Gas Save value Change value Alarm Display When A1 alarm occurs and the operator recognizes it and presses Power key only th...

Page 14: ...anged LEL is displayed at Calibration Mode In the event that VOL Alarm Value is changed VOL is indicated at Calibration Mode If you don t want the display by VOL move to the next step without changing...

Page 15: ...ment and components as below Instrument driver wrench hexagon Battery Energizer No E91 EN91 or GP rechargeable Battery GP210AAHCB Sensors for replacement Filters for replacement Disassembly 1 1 2 3 4...

Page 16: ...permitted to use other batteries for the instrument at dangerous regions Caution 1 It is absolutely prohibited to replace battery at potential explosion or dangerous regions 2 Replacement of componen...

Page 17: ...nnecting the pump to the instrument Prior to use make sure that the instrument is tightly attached to the probe cover which is connected to the sensor Please note that External Pump is the optional pr...

Page 18: ...Operation Flow Chart 17...

Page 19: ...18 Operation Flow Chart Calibration Configuration Flow...

Page 20: ...performed at the same or similar environment as that of the instrument use temperature pressure and humidity If temperature changes sharply during use of the instrument for instance using the instrum...

Page 21: ...Certificates 20...