CamCor CT and PRO Series Hardware Manual
103
Section 8
H/L Alarm Output
Assigning the high/low alarm output to the status output allows a high/low alarm error to be output. Below are the available
parameters that can be configured as high/low alarm outputs:
•
Mass Flow:
Instant mass flow rate
•
Vol Flow:
Instant volume flow rate
•
Dens:
Density
•
Temp:
Temperature
•
Counter 1:
Counter 1
•
Counter 2:
Counter 2
Configuring a High/Low Alarm Output
1. Select “H/L Alarm” as the status output function assignment (see
, No. 2-4-5-1).
2. Configure the active output (see
, No. 2-4-5-3). The following outputs become available
based on the selections made above:
–
N.C.:
Output is Closed when the preset state is reached.
–
N.O.:
Output is Open when the preset state is reached.
When the settings above take effect, the functions are ready for use.
Drive Output Alarm (for Maintenance Purposes)
Assigning the drive output alarm to the status output allows an alarm to be activated when the drive output exceeds the us
-
er-established threshold.
Configuring a Drive Output Alarm
1. Select “Drive O Alm.” as the status output function assignment (see
2. Enter the threshold to indicate at what percentage of the drive output the alarm should be activated. The default is
100%. See
, No. 2-4-5-3 for more information.
3. Configure the active output (see
, No. 2-4-5-3). The following outputs become available
based on the selections made above:
–
N.C.:
Output is Closed when the preset state is reached.
–
N.O.:
Output is Open when the preset state is reached.
When the settings above take effect, this function is enabled.
No Function/Assignment
When “no function” is selected as the status output function assignment, the following status output appears:
–
N.C.:
Output is always Closed.
–
N.O.:
Output is always Open.
Configuring a No Function Output
, No. 2-4-5-1).
Summary of Contents for CamCor CC001
Page 14: ...14 CamCor CT and PRO Series Hardware Manual Section 3 This page is left blank intentionally ...
Page 24: ...24 CamCor CT and PRO Series Hardware Manual Section 3 This page is left blank intentionally ...
Page 54: ...54 CamCor CT and PRO Series Hardware Manual Section 5 This page is left blank intentionally ...
Page 62: ...62 CamCor CT and PRO Series Hardware Manual Section 7 This page is left blank intentionally ...