48
CamCor CT and PRO Series Hardware Manual
Section 5
Note
Vibrating copper tubing caused by the flow meter in operation can produce zero shifts. Increase the number of fixed points
to reduce vibration.
To minimize transmission of oscillation to the sensor, the spirally wound copper tubing should start and end at the manifold
of flanged sensor housing.
Sudden temperature change may damage the performance of the flow meter. Keep the temperature change of the fluid
within ±53.6°F/min. for both heating and cooling.
CAUTION
Do not use electric heaters which could produce noise. If ambient temperature around the transmitter is expected
to exceed 131°F, a separately-mounted transmitter is recommended.
Copper
tubing
Copper
tubing
Spot
weld
Piping
Steam
Pipe retainer
(material SUS316L strip)
CC00A, CC001
CC003
Sensor unit
Sensor
unit
Figure 5.17—Heat and cold retention via copper tubing and pipe retainers (CT Series CC00A, CC001, and CC003 meters)
Do not apply heat retention to
terminal box and mounting tube.
Apply heat retention
only to the main body.
0.79 in.
minimum
Spot
weld
Piping
Pipe retainer
(material SUS316L strip)
Manifold
Manifold
Copper
tubing
Transmitter
CC006 through CC080
CC025 through CC080
High- and Low-temperature Models
Sensor unit
Figure 5.18—Heat and cold retention via copper tubing and pipe retainers (CT Series CC006 to CC080 meters)
Summary of Contents for CamCor CC001
Page 14: ...14 CamCor CT and PRO Series Hardware Manual Section 3 This page is left blank intentionally ...
Page 24: ...24 CamCor CT and PRO Series Hardware Manual Section 3 This page is left blank intentionally ...
Page 54: ...54 CamCor CT and PRO Series Hardware Manual Section 5 This page is left blank intentionally ...
Page 62: ...62 CamCor CT and PRO Series Hardware Manual Section 7 This page is left blank intentionally ...