Calibration and Testing
Register
(Decimal)
Register
(Hex)
Description
Default
Data
Type
Data Range
Access
157
009D
Zero Offset
0
FP
– 999999 to 999999
RW
178
00B2
Trigger Self Diagnosis
U8
0 = Self Diag. Mode 1
1 = Self Diag. Mode 2
2 = Self Diag. Mode 3
3 = Self Diag. Mode 4
4 = Self Diag. Mode 5
RW
179
00B3
LCD Test
U8
1 = Backlight Test
2 = LED Test
3 = LCD Test
RW
180
00B4
Trigger Auto Zero Offset
U8
Any data will start auto
zero offset
RW
204
00CC
Trigger for High-Density Calibration
U32
RW
Holding Registers
Register
(Decimal)
Register
(Hex)
Description
Default
Data
Type
Data Range
Access
227
00E3
Mass Flow Rate Unit
see
U8
RO
228
00E4
Mass Flow Rate Reading
FP
230
00E6
Volume Flow Rate Unit
U8
231
00E7
Volume Flow Rate Reading
FP
233
00E9
Density Unit
U8
234
00EA
Density Reading
FP
236
00EC
Internal Temperature Unit
U8
237
00ED
Internal Temperature Reading
FP
239
00EF
Output Saturation Alarm
see
U8
RO
240
00F0
Sensor Failure Alarm
241
00F1
Transmitter Failure Alarm
242
00F2
Parameter Error Alarm
243
00F3
Calibration Failure Alarm
244
00F4
Slug Flow Alarm
245
00F5
Transmitter Alarm
246
00F6
Output Override Status
247
00F7
Calibration in Progress
248
00F8
High/Low Alarm
249
00F9
Maintenance in Progress
250
00FA
Transmitter Operation
251–255
Reserved
256
0100
Analog Output 1 Output (%)
FP
RO
258
0102
Analog Output 2 Output (%)
260
0104
Status Input Status
U8
RO
261-267
Reserved
18
Section 3
CamCor
™
Modbus Protocol
Summary of Contents for CamCor
Page 1: ...CamCor Modbus Protocol Manual Manual No 50284150 Rev 01 ...
Page 4: ...This page is left blank intentionally iv Table of Contents CamCor Modbus Protocol ...
Page 8: ...This page is left blank intentionally 8 Section 2 CamCor Modbus Protocol ...
Page 10: ...This page is left blank intentionally 10 Section 2 CamCor Modbus Protocol ...
Page 22: ...This page is left blank intentionally 22 Section 3 CamCor Modbus Protocol ...