Pag. 11

3.  Menu

The calibration keypad has a menu with several options, each one of them allows to per-

form different actions on the detector. In the following table are summarised all the possible 

actions.

Table 2)  Calibration keypad menu description

D i s p l a y e d 

name

Meaning

Description

PSd

Password

1234. It allows to access the keypad menu, otherwise the user 

can only reset the detector or visualise its  firmware version. 

To modify the alarm thresholds and the hysteresis value, the 

password is 5345.

ZEro

Zero

When the detector is turned on for the first time, or when 

a new kit sensor head is mounted into the instrument, the 

ZERO calibration is recommended.

This action has to be done under the following two condi-

tions:

-The detector is to be in fresh air (without any gas or interfe-

ring compounds);

-it must be powered on at least for 8 hours.

Span

Span

The Span allows the instrument calibration and is to be exe-

cuted if the response in gas of the instrument is not correct.  

To run the Span calibration, it is essential the use of a sample 

gas bottle with Target Gas.

If the concentration of the sample gas bottle is different from 

that displayed on the keypad, then must be performed the 

Span calibration. Otherwise this step is not necessary.

For the correct calibration of gas detectors, Sensitron makes 

available a dedicated calibration case complete with all ne-

cessary tools to perform accurate tests. The calibration KIT, 

supplied with a carrying case, comprises the necessary in-

strumentation to verify in situ the response of Sensitron's 

gas detectors.

The calibration cap can be fitted on all of Sensitron sensing 

heads by employing, when necessary, the endowed adap-

tors. Both gaskets and screw threads inside the cap allow an 

optimum tightness between the sensing head and the cap. 

The analysis chamber inside the calibration cap has been de-

signed and realized to obtain a test method identical to the 

one adopted in our Laboratory for the initial factory calibra-

tion.  The flow meter valve is to be mounted on the gas can 

to allow the gas flow at a controlled velocity to obtain accu-

rate data to verify the sensing element status. 

t4.20

4-20 cali-

bration

4-20 mA output calibration

Summary of Contents for STS/CKD+

Page 1: ...Manuale d istruzione Instruction Manual MT4508 rev 0 01 10 2020 2020 SENSITRON S R L All rights reserved www sensitron it Tastierino di calibrazione Calibration keypad STS CKD...

Page 2: ...for any additional information relative to use or repair QUESTO MANUALE DEVE ESSERE LETTO ATTENTAMENTE DATUTTI COLORO CHE HANNO O AVRANNO LA RESPONSABILITA DI INSTALLARE UTILIZZARE O DI PRESTARE UN S...

Page 3: ...rors 15 5 Troubleshoot of detector s errors 15 6 Warranty 16 7 Instructions for disposal 16 SOMMARIO 1 Introduzione 1 1 1 Descrizione generale 1 2 Connessione e autotest 2 3 Menu 3 4 Schemi di funzion...

Page 4: ...ativa isteresi Reset del dispositivo Head pairing accoppiamento del rivelatore con la testa sensore Modifica o sola visualizzazione del Baud Rate Impostazione dell indirizzo del bus RS485 Visualizzazi...

Page 5: ...cendono Quando il rilevatore esce dalla fase di warm up la scritta HEAT scompare e rimane acceso solo il LED ON Una volta collegato il tastierino eseguir un autotest interno Questo passaggio permette...

Page 6: ...ratura dello Span permette la modifica della sensibilit al gas del rilevatore Per la calibrazione dello Span sensibilit al gas richiesta una bombola di gas campione Se la concentrazione di gas della b...

Page 7: ...ione Le voci disponibili per ogni classe di rilevatore sono elencate nella tabella sottostante Tabella 3 Voci del menu disponibili per ogni tipologia di rilevatori SMART SOTTO MENU Tipologia rivelator...

Page 8: ...ER ESC ENTER ENTER PWd OK La cifra in modi ca lampeggiante Set Value Set Digit Set Digit ENTER Set Digit ENTER Set Digit ENTER ENTER ESC ESC ESC ESC ENTER ENTER ENTER Start ENTER ESC ENTER ENTER ESC 4...

Page 9: ...NTER ENTER ENTER ESC ENTER Start ENTER ENTER Start ENTER ENTER La cifra in modi ca lampeggiante Set Value Set Digit Set Digit ENTER Set Digit ENTER Set Digit ENTER ENTER ESC ESC ESC ESC ENTER ESC ENTE...

Page 10: ...di impostazione del baudrate ErrB Inserimento di un valo re non ammesso possi bili valori 9600 19200 38400 115200 Non viene memorizzato il valore inserito 5 Risoluzione dei problemi del rilevatore Ta...

Page 11: ...di questa pubblicazione pu essere riprodotta distribuita o trasmessa in qualsiasi forma o con qualsiasi mezzo inclusi fotocopie registrazioni o altri metodi elettronici o meccanici senza la previa aut...

Page 12: ...teresis To reset the device To couple detector with the sensor head Head pairing To change or visualize only the Baud Rate To set the address of the RS485 bus To show the firmware version of the detec...

Page 13: ...while the detector is warming up the HEAT message appe ars and all the LEDs light up When the warm up phase ends the HEAT message disappears and only the ON LED remains on Once connected the calibrat...

Page 14: ...pan calibration it is essential the use of a sample gas bottle with Target Gas If the concentration of the sample gas bottle is different from that displayed on the keypad then must be performed the S...

Page 15: ...15200 Depending on the detectors series different menu entries are available The following table reports the available items per each type of detector Table 3 Menu component available per each detecto...

Page 16: ...ER ENTER ESC ESC ESC ESC ENTER ESC ENTER ENTER ENTER ESC ENTER ENTER ENTER ESC ENTER ENTER ENTER ESC ENTER ESC ENTER ESC ENTER ENTER ESC ENTER ENTER ESC ENTER ENTER ESC ENTER ESC ENTER ESC ENTER ENTER...

Page 17: ...git Set Digit ENTER Set Digit ENTER Set Digit ENTER ENTER ESC ESC ESC ESC The digit blinks when the user edits it ENTER ENTER ENTER ENTER ENTER ESC ENTER Start ENTER ENTER Start ENTER ENTER ENTER ESC...

Page 18: ...t a value out of range range 1 247 The address is not memo rised Baudrate error ErrB The user has set a non ad missible value admissible values 9600 19200 38400 115200 The baudrate value is not memori...

Page 19: ...Warranty Warranty on Sensitron products is valid one year from the manufacturing date placed on the product 7 Instructions for disposal When the device reaches the end of its life it should be dispos...

Page 20: ...actdetails SENSITRONs r l Viale della Repubblica 48 20007 Cornaredo MI Italy TEL 39 0293548155 FAX 39 0293548089 www sensitron it E mail sales sensitron it LEADER IN SIL2 SIL3 FIXED GAS DETECTION SYST...

Reviews: