6.
Initial Startup
When power is applied to the SEC3120, it enters its power up sequence (approximately sixty seconds),
followed by the sensor warm-up mode which can take between one and five minutes depending on the
sensor(s). At the end of the warm-up period with no faults present, the SEC3120 automatically enters the
normal operating mode.
If a fault is present after warm-up, the LCD and the Fault LED will indicate the fault. See your specific
sensor’s manual for warm-up and fault current value meanings (less than 4 mA).
NOTE:
If your SEC3120 unit is set to
operate two sensors in the ‘Identical’ Role and they are not identical
in gas type, range or units, then the SEC3120 unit will display a fault and indicate that the
Sensors Are Not
Identical
, and the fault relay on the SEC3120 unit will be engaged.
Inital Power Up Screen Sequence of the SEC3120
‘BRWN’ – Brown Out Reset (or ‘SOFT RESET’ shown
on bottom line for a software reset)
‘005’ – Counts down to ‘000’ second startup delay
‘VER’ refers to the Firmware version installed in the
display.
‘INIT ROM FROM RAM’ – permanent parameters
transferred from either ‘ROM’ to NVRAM or from
N
VRAM to ROM, or ‘ROM’ and ‘RAM’ are ‘IDENTICAL’.
‘BOOT SENSOR SYSTEM’ – Starting up sensor
controllers
Gas Sensor Port (1
x
¾NPT
)
SEC 3000 / 3300
SEC Millenium / Signature / 5000
Magnetic Switch (3)
Field Wiring Port (2
x
¾NPT
)
Power
2x 4-20 mA
RS485 (MODBUS) A & B
Alarm and Fault Relays
Fault LED
Green
OK
Red
Fault
LCD Display:
Real Time Status
Gas Type
Gas Concentration
Fault Identification
Range
ID Number
Memory Card Status
Interactive Menus
Calibration
Relay Programming
ID & Zone Number
Mounting holes
(2)
Memory Data Card Slot
Alarm LED
Green
OK
Yellow
Low Alarm
Orange
Mid Alarm
Red
High Alarm
SEC3120 TRANSMITTER
-STARTING UP FROM
BRWN
(SOFT RESET)
005
SEC3120 TRANSMITTER
VER X.YY.RRRR
005
08/10/2010
21:15:33
SEC3120 TRANSMITTER
VER X.YY.RRRR
003
-INIT ROM FROM RAM
SEC3120 TRANSMITTER
VER X.YY.RRRR
-BOOT SENSOR SYSTEM