4
›
english
-20
-LT
-MT
-HT
-XT
-CT
0
100
Fixed (e.g. -MT: 0°C @ 4 mA, 250°C @ 20 mA)
-IT models: 4 to 20 mA output configurable within this range
-MBT and -MB models: Digital output, full temperature range
250
500
1000
MEASUREMENT TEMPERATURE RANGE (°C)
OUTPUT AND INTERFACE
-I
4 to 20 mA output, no touch screen
-IT
4 to 20 mA output and two alarm relay outputs, with touch screen
-MB
RS485 Modbus output, no touch screen
-MBT
RS485 Modbus output and two alarm relay outputs, with touch screen
Diameter of target
spot measured versus
distance from sensing
head at 90% energy
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
11.9
45.2
78.6
0.5
1.8
3.1
0
19.7
39.4
0
11.9
0
500
1000
Spot Dia.
(inches)
Spot Dia.
(mm)
Spot Dia.
(inches)
Spot Dia.
(mm)
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
61.9
111.9
0.5
2.4
4.4
0
4
8
100
200
D:S 15:1
D:S 2:1
-102
-115
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
11.9
36.9
61.9
0.5
1.5
2.5
0
19.7
39.4
0
500
1000
Spot Dia.
(inches)
Spot Dia.
(mm)
D:S 20:1
-120
Spot Dia.
(inches)
Spot Dia.
(mm)
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
11.9
0.5
0
0
5.0
0.20
100
3.9
12.5
0.49
200
7.9
-CF5
Distance: Sensor to object (inches)
Distance: Sensor to object (mm)
11.9
28.6
45.2
0.5
1.1
1.8
0
19.7
39.4
0
500
1000
Spot Dia.
(inches)
Spot Dia.
(mm)
D:S 30:1
-130
FIELD OF VIEW
SENSING HEAD OPERATING TEMPERATURE RANGE
-MT
0°C to 60°C
-HT
0°C to 180°C
The high ambient temperature sensing head on -HT models is capable of withstanding temperatures
of up to 180°C without cooling. It is available with 20:1 optics.
There is no need to supply cooling air or water, and the miniature sensing head is much smaller than
bulky, cooled sensors.