Congratulations on your purchase of your Sensorex SD7000 

Series pH or ORP true differential electrode.  Please follow 

instructions and tips to get the maximum life out of your 

electrode.

Introduction

Electrode Care and Use Tips

1. The SD7000 Series pH and ORP electrodes are shipped with a  

    cap containing a solution of pH 4 buffer and potassium  

     

    chloride.  

    The cap should remain on the electrode until it is used. If the 

    electrode is used infrequently the cap and its solution should 

    be saved and the electrode stored in it. Store pH and ORP 

    electrodes in pH 4 buffer if soaking solution is discarded. 

    Before using the electrode remove the soaker cap as shown     

   in FIG 1.

2. Sensorex true differential electrodes have a 2 year prorated           

warranty from the date etched in the electrode body as shown         

   in FIG 2.   Electrodes in inventory should be rotated so that the      

    older electrodes are used first. The date code is stamped on 

each electrode by fiscal week and year (Example: 4613 = week 2    

of November 2013). 

3. After exposure to a sample, buffer or rinse solution, carryover 

    can be minimized by blotting—never by wiping—the elec-

    trode with a clean non-abrasive paper or a clean cloth towel. 

    

Do not use a brush on pH glass. 

4. As a rinse solution, measure and use a portion of the next 

    sample or buffer. This action will minimize carryover 

    contamination. 

5. When calibrating pH electrodes, use a buffer close in value to 

    that expected from the sample for one-point calibrations or as 

    the second buffer for two-point calibrations (See below). This 

    action will minimize span errors. ORP electrodes are typically 

    not calibrated. However, a solution to verify electrode perfor-

    mance should be used periodically. 225mV solution, Zobell's 

    solution (Part no. B225), is a good choice for single point 
    verification.

                                                                                                             

 PRODUCT INSTRUCTION SHEET 

Parts covered by this product data sheet include: 

SD7000CD, SD7000CD-ORP, SD7500CD, SD7500CD-ORP, SD7420CD, SD7420CD-ORP

    SD7000 Series Differential 
  pH/ORP Electrode Instructions

 FIG. 1

 FIG. 2

Form:  InstrSD7000-05 [Rev: 2016-08-15] 

©2016 Sensorex Corporation

 

Mo

de

l S

D7

00

0C

D D

iffe

ren

tia

l p

H

4615

Summary of Contents for SD7000 Series

Page 1: ...iscal week and year Example 4613 week 2 of November 2013 3 After exposure to a sample buffer or rinse solution carryover can be minimized by blotting never by wiping the elec trode with a clean non ab...

Page 2: ...span errors but will not affect the speed of response 8 Salt bridge should be replaced when the electrode readings cannot be corrected by the meter s controls and or when their speed of response is t...

Page 3: ...lean warm water If the sensor is usually in a process above 7 pH it is advisable to soak it 4 5 minutes in a weak acid solution hydrochloric recommended Place the sensor back in the soap solution for...

Page 4: ...ibration must be performed ORP is usually not calibrated but quinhydrone based solution or other standards can be used to verify ORP sensor performance See FIG 9 Two Point Calibrations Two point calib...

Page 5: ...isconnected from un powered meters Electrode Wiring See the figures 10 through 17 to find wiring for your particu lar pH or ORP controller or transmitter Figures 10 17 apply to SD7000CD and SD7000CD O...

Page 6: ...ard cell solution 5 With the multimeter still connected the same way rinse the sensor with water and put it in a pH 4 or pH 10 buffer solution Wait for the temperature of the sensor and buffer to reac...

Page 7: ...ed leads 1 5 NPT mounting self powered Replacement salt bridges sold separately SD7500CD ORP Differential ORP sensor with replaceable salt bridge 20ft cable tinned leads 1 5 NPT mounting self powered...

Page 8: ...x for details SENSOREX offers no warranty either expressed or implied as to the useful life of these products There are no implied warranties of merchantability or fitness for a particular purpose giv...

Reviews: