background image

                                                                                                             

 PRODUCT INSTRUCTION SHEET 

Page 4 of 8

Form:  InstrSD7000-05 [Rev: 2016-08-15] 

©2016 Sensorex Corporation

Electrode Calibration Guidlines:

As a rule, follow the procedures shown in the pH Meter's 

Instruction Manual. These procedures will vary depending on 

whether the meter is a simple type with manual adjustments, a 

micro-processor type or a pH transmitter.

The frequency of calibration is a function of many factors.  These 

factors include:

1) The accuracy required by the application.

2) The value of the off-specification product versus the cost  

    of calibration.

3) The coating or abrasive nature of the application.

4) The stability of the pH Electrode and pH Meter as a system.

The frequency of calibration is really determined by experience. 

At a new installation, calibration might initially be checked every 

few hours or shift with the calibration changes

noted in a log. As a pattern of longer stability is found, the time 

between calibration checks can be increased to once a day or 

once a week.  

System Calibration Concepts

The pH Electrode and the pH controller should always be cali-

brated as a system. Electronic calibration of a pH controller with a 

pH signal simulator checks the controller only and does not cor-

rect for imperfections of the pH electrode. Even if perfect when 

new, the performance of pH electrodes varies with time, usually 

in an unpredictable way. When changing electrodes or connect-

ing an electrode to a different pH controller, re-calibration must 

be performed. ORP is usually not calibrated but quinhydrone 

based solution or other standards

can be used to verify ORP sensor performance.  See FIG. 9.

Two-Point Calibrations

Two-point calibrations correct for both the pH electrode's offset 

and span errors. Since both the offset and span vary with time 

the two-point method is the onr preferred. Choose buffer pH 7 

for zero-point and a second buffer close to your normal operat-

ing range (usually pH4.01 or pH 10.00).  See FIG. 8 A-C. 

Grab Sample Calibrations

The Grab Sample Calibration method is used when it is difficult 

or undesirable to remove an electrode from a system.  This meth-

od involves obtaining a sample of the liquid being measured and 

noting the meter's reading at that time. The sample's reading is 

obtained by use of a calibrated lab or portable meter and that 

reading is compared to that of the on-line meter. The on-line 

meter is adjusted by the difference between the readings. It is 

important to use the difference between the readings because 

the system's reading may have changed in the intervening time. 

It is important that the sample being measured by the lab meter 

be at the process temperature or erroneous results may occur. 

  pH

  4.0

  BU

FFE

R

 pH 4.01 

 BUFFER

pH Meter

7.00pH

pH 4.01/10.00

pH 7.00

 FIG. 8A

 FIG. 8B

 FIG. 8C

 

PLATINUM ORP ELECTRODE IN 7 BUFFER/QUINHYDRONE MIXTURE

 

 

Temperature 

    20C (68F) 

       25C (77F) 

      30C (86F) 

Readings (mV) 

   89-107 

      83-101 

      76-94 

Readings (pH) 

   5.20-5.50 

      5.30-5.60 

     5.42-5.72 

 

PLATINUM ORP ELECTRODE IN 4 BUFFER/QUINHYDRONE MIXTURE

 

 

Temperature 

    20C (68F) 

       25C (77F) 

      30C (86F) 

Readings (mV) 

   260-287 

      254-281 

      247-274 

Readings (pH) 

   2.15-2.60 

      2.25-2.70 

     2.37-2.82 

 

 FIG. 9

Summary of Contents for SD7000 Series

Page 1: ...iscal week and year Example 4613 week 2 of November 2013 3 After exposure to a sample buffer or rinse solution carryover can be minimized by blotting never by wiping the elec trode with a clean non ab...

Page 2: ...span errors but will not affect the speed of response 8 Salt bridge should be replaced when the electrode readings cannot be corrected by the meter s controls and or when their speed of response is t...

Page 3: ...lean warm water If the sensor is usually in a process above 7 pH it is advisable to soak it 4 5 minutes in a weak acid solution hydrochloric recommended Place the sensor back in the soap solution for...

Page 4: ...ibration must be performed ORP is usually not calibrated but quinhydrone based solution or other standards can be used to verify ORP sensor performance See FIG 9 Two Point Calibrations Two point calib...

Page 5: ...isconnected from un powered meters Electrode Wiring See the figures 10 through 17 to find wiring for your particu lar pH or ORP controller or transmitter Figures 10 17 apply to SD7000CD and SD7000CD O...

Page 6: ...ard cell solution 5 With the multimeter still connected the same way rinse the sensor with water and put it in a pH 4 or pH 10 buffer solution Wait for the temperature of the sensor and buffer to reac...

Page 7: ...ed leads 1 5 NPT mounting self powered Replacement salt bridges sold separately SD7500CD ORP Differential ORP sensor with replaceable salt bridge 20ft cable tinned leads 1 5 NPT mounting self powered...

Page 8: ...x for details SENSOREX offers no warranty either expressed or implied as to the useful life of these products There are no implied warranties of merchantability or fitness for a particular purpose giv...

Reviews: