Page 18
Oryx DTS User Manual v4
The “Edit Measurement Configuration” button enables the user to set an active
configuration. The default_idle configuration can be modified to start taking
measurements although the user is recommended to enter a new installation name.
Enable channels using the “Channel specific settings” tab and the “Channel details”
sub-tab. Measurement or signal averaging time can be set, as well as the overall
repetition time under the appropriate tabs. For example, setting 2 channels active,
measuring each channel for 5 minutes, repetition time 1 hour, would result in 5
minutes measurement for each channel repeated every hour. Select the
“Measurement Config Wizard” button to be guided through the measurement setup
process of the DTS.
NOTE: If the idle time is greater than 60 seconds, the laser and data acquisition
module will be powered down between measurements to reduce power
consumption.
Please see section 0 for how to set up an Oryx with the ODC and a description of
creating and editing configuration files and section 0 for a guide to the configuration
wizard.
NOTE: Clicking “UPLOAD” will upload a new configuration to the Oryx and will
DELETE ALL MEASURMENTS on-board the Oryx at that time. (Any data already
downloaded to the PC is unaffected).
If selecting to start measurements as soon as possible, there may be a delay of up
to one repetition time before commencement of the first measurement.
Once the Oryx has been set up with a measurement configuration, click the “Collect
data” button to begin monitoring each Oryx connected to the base station PC. The
ODC software will receive a “data_ready” signal from each Oryx when a
measurement has been recorded and will automatically download all data from the
DTS. The user can also query the status of each Oryx by clicking the “Get Status
Update” button on the status panel (see Figure 25).
NOTE: Data will only be downloaded from the Oryx when the ODC is running and
the “Process Log” or “Oryx Status Panel” screen is visible. If this software is not
running or there is a communications failure, data will be stored on the Oryx until
communication is restored.
NOTE: To avoid frequent communications traffic with the Oryx, the ODC software
will wait for a “data_ready” message from the Oryx before downloading data. The
user can force all data to be downloaded by simply clicking the relevant “Get Status
Update” button on the Oryx status panel.
Summary of Contents for ORYX DTS
Page 1: ...This page intentionally blank ...
Page 2: ......
Page 5: ......
Page 6: ...This page intentionally blank ...
Page 70: ...Page 64 Oryx DTS User Manual v4 Figure 45 Setting a zone Figure 46 The list of defined zones ...
Page 100: ...Page 94 Oryx DTS User Manual v4 This page intentionally blank ...
Page 101: ......