Page 62
Oryx DTS User Manual v4
Figure 43 - Offset correction – point reference type selected
Setting the calculation method
Once the offset calibration is complete, if any of the calibration options have been
set as ‘matched sections’ or ‘reference section’ the user will be presented with a
screen to set the method of the calibration calculation (Figure 44). The two options
are ‘Dynamic’ and ‘Fixed’ and can be set independently for slope and offset
calibration.
•
Dynamic – in this case the calibration coefficient used will be calculated
every measurement. Hence the coefficient may change from trace to trace.
•
Fixed – in this case the calibration coefficient calculated during the setup
wizard will be fixed and used for all calculations of temperature data traces
In most cases, it is recommended to set the method to ‘fixed’ so that the calibration
coefficients will not change from measurement to measurement. However, in
certain cases, for example in harsh environments where the fibre properties may
change, dynamic method may be appropriate. (In this case however a ‘combined’ or
double-ended measurement may be appropriate).