Oryx DTS User Manual v4
Page 89
SPECIFICATIONS
PERFORMANCE SPECIFICATIONS
Performance measured with the Oryx DTS at 25°C ambi ent temperature unless
otherwise specified. Sensornet Ltd reserves the right to modify these specifications
at any time.
Parameter
Unit
Value
Comments
SR
XR
Measurement time
s
≥
15
≥
30
No upper limit
Temperature accuracy
°C
+/-0.5
Note 1, 3
Max optical loss
dB
7
4.5
Note 2
Maximum Range
km
5.3
12
Max fibre length
km
12.5
Note 2a
Temperature
resolution
°C
Refer to
www.sensornet.co.uk
for latest spec.
Temperature range
°C
-200 to 600
Note 5
Sampling period
M
1
Spatial Resolution
M
2
Note 4
PT100 temperature
range
°C
-50 to +250°C
PT100 accuracy
°C
+/-0.25
-200 to +200°C; Note 6
RTC accuracy
Min/Yr
+/-4
-40 to +65°C
Note 7
Battery lifetime
Yr
15
Note 8
Fibre type
50/125
µ
m graded index
multimode
Note 9
Summary of Contents for ORYX DTS