MULTISENS User Manual
11
MultiSens User Guide EN_v1.0_281218
4.
Event Log
Up to 30 events may be saved, the oldest data will be automatically substituted with the newest one. The saved
data can be downloaded to PC through SENSOTRAN IR-LINK.
Data log records the operation status every second for more than 2 months.
Log Categories
Log Details
EVENT(High, Low, TWA, STEL)
Alarm
Time, Duration, Alarm Type, Gas Concentration, Serial Number
BUMP TEST Log
Test date, Pass/non-pass, Span Gas Concentration, Measured Concentration
Calibration Log
Calibration Date, Type, Span Gas Concentration, Measured Concentration
Data Log
Time, Date, Concentration, Alarm Type, Options
5.
Calibration
<Caution> The initial calibration is executed at SENSOTRAN S.L. before device shipment. The calibration
values are saved in the device which means inaccurate calibration may impair the accuracy of the device
performance. User should perform calibration every six months.
<Caution> Because it is calibrated on the assumption that oxygen concentration is 20.9%vol, the combustible
gas is 0%LEL, and the toxic is 0ppm in the normal fresh atmosphere, fresh air calibration must be conducted
in the absolutely clear air without any impact of other gases.
Fresh air calibration in the airtight spaces
therefore is not recommended. Make sure to avoid calibration under the work environment where gases may
be present.
5.1.
Fresh Air Calibration
Press KEY button ( ) for 3 seconds under the gas calibration value mode and the icon ( ) for fresh air
calibration will appear on the display with the message “CAL ZERO.” Press for another 3 seconds to do fresh air
calibration and it will take 10 seconds to calibrate. Press the button during the calibration process to abort the
calibration. If you press the button upon the completion, it will return to the fresh air calibration mode, and if
you don’t press the button, it automatically enters the measure mode.
If the calibration fails, message FA(Fail), appears on the display. Press the button to enter
the initial fresh air calibration mode and it changes into the measure mode if you do not press
the button for 3 seconds. If FA continues, please consult SENSOTRAN or your local
distributor as it may require the replacement of the sensor or maintenance tasks.
3 sec.
3 sec.
10 seconds countdown
(ZERO ADC currently displayed)