background image

Warning:

 The measurement of SpO

2

 and PR with SenTec TC Sensors is 

only defined on sites specified by the manufacturer (refer to the instrument`s 

manual). In order to avoid erroneous readings and false alarms of SpO

2

 and 

PR, ensure that the appropriate patient type (Adult) is selected. Ensure to 

disable the parameters SpO

2

/PR for sensor application on other measurement 

sites.

Warning:

 The use of temperatures higher than 41°C requires special 

attention to patients with susceptible skin, e.g. neonates, geriatric patients, 

burn victims, patients with skin diseases. 

CAUTION: 

A flat, well-perfused area of intact skin is needed (centrally 

located sites are preferable). Avoid placement over large superficial veins or 

areas of skin breakdown or edema.

Cleaning and Disinfection

CAUTION: 

The use of other cleaning and disinfection procedures oth-

er than those recommended here may result in permanent damage to the 

sensor.

CAUTION: 

Do not remove the membrane of the sensor for cleaning 

and disinfection.

CAUTION: 

Wiping and/or the application of mechanical force in general 

can damage the sensor membrane or the sensor cable mechanically. Wipe the 

sensor carefully and inspect the condition of the membrane and sensor cable 

after cleaning and/or disinfection.

The OxiVenT™ Sensor must be cleaned before inserting it into the docking 

station in order to remove any gel residues or other soils such as blood, protein, 

or mucous that could contaminate the docking station, cause cross-infections 

between different patients and/or gas leakage during calibration. 
In order to remove all foreign substances, the sensor surface should be cleaned 

prior to disinfection.

Important: 

The cleaning and disinfection solutions listed here have been 

tested for their suitability for the OxiVenT

TM

 Sensor. The below listed name brand 

products for cleaning and disinfection may be substituted by other name brand 

products of equivalent composition. However, as there are many factors (contact 

time, force applied during wiping etc.) that can influence the functionality of the 

sensor, it is the responsibility of the user to test the functionality of the sensor 

after cleaning and/or disinfection. SenTec recommends using 70% Isopropanol.

Table 1: Low Level Disinfection

Brand name/

manufacturer

Cleaning procedure

WEBCOL Alcohol Prep 

Pads

 (Kendal); 70% 

Isopropanol

Carefully wipe all SenTec TC Sensor surfaces and 

the cable with alcohol wipes and allow to dry.

Mikrozid

 (Schülke Mayr) 

pre-soaked wipes; 

25% Ethanol, 35% 

Isopropanol

Kodan® forte 

(Schülke&Mayr);

45,0 g 2-Propanol, 10,0 g 

1-Propanol, 0,20 g Biphe-

nyl-2-ol, 30% H2O2

Carefully wipe all surfaces of the SenTec TC Sen-

sor and the cable with wipes soaked Kodan forte. 

Wait 5 min. Remove all Kodan forte residues using 

70% Isopropanol. Allow to dry.

Gigasept® AF

 

(Schülke&Mayr); 

15 g Didecyldimethylam-

moniumchlorid,

10 g Phenoxypropanole, 

6,9 g Aminoalkylglycine; 15 

– 30% non-ionic tensides.

Carefully wipe all surfaces of the SenTec TC 

Sensor and the cable with wipes soaked in 4 % 

solution (e.g. 960 ml deionoized water plus 40 

ml Gigasept AF). Note that the 4% dilution has a 

shelf-life of 7 days. Wait 15 min. 

Remove all Gigasept residues using 70% Isopro-

panol. Allow to dry.

OxiVenT™ Sensor

 

REF OV-A/P/N

Carbon Dioxide Tension, Oxygen Tension and 

Oxygen Saturation Sensor
Reusable / Nonsterile / Waterproof

Directions for Use

Introduction

SenTec’s OxiVenT™ Sensor is intended for continuous PCO

2

, PO

2

, SpO

2

/PR 

monitoring.
The OxiVenT™ Sensor provides superior performance, is robust, reliable and 

requires comparatively low maintenance. It combines within a patented digital 

sensor design the optical components needed for 2-wavelength, reflectance 

pulse oximetry with the components needed to measure PCO

2

 and PO

2

.

Indications/Contraindications

The OxiVenT™ Sensor (REF OV-A/P/N) is indicated for use when continuous 

non-invasive monitoring of carbon dioxide tension, oxygen tension, oxygen 

saturation, and pulse rate are required for adult and pediatric patients. In 

neonatal patients the use of the OxiVenT™ Sensor is indicated for carbon 

dioxide tension and oxygen tension monitoring only. Oxygen tension monitoring 

is contraindicated for patients under gas anesthesia. The OxiVenT™ Sensor 

is for use only with SenTec instruments and instruments that contain SenTec 

monitoring technology or are licenced to use SenTec sensors (SenTec-

compatible instruments).

Instructions for use

Preparation - Membrane change

The membrane and the electrolyte of the OxiVenT™ Sensor must be changed 

regularly. The monitor requests a membrane change if required. Additionally, 

the membrane of the sensor must be changed if it is damaged, does not have 

a tight fit, or if there is trapped air or dry electrolyte under the membrane. For 

further information refer to the Instruction Manual for the SDMS, the Membrane 

Changer Directions for Use for detailed membrane change instructions, and the 

Technical Manual for the SDM or contact the manufacturer of the instrument.

Preparation - Calibration

The PCO

2

 segment and - less frequent - the PO

2

 segment of the OxiVenT™ 

Sensor must be calibrated regularly. Between monitoring it is recommended to 

store the sensor in the docking station. This will ensure that the sensor is ready 

for use, as the sensor will automatically be calibrated as needed. For further 

information refer to the Instruction Manual for the SDMS and the Technical 

Manual for the SDM or contact the manufacturer of the instrument.

Patient Application of the OxiVenT™ Sensor 

The choice of the sensor attachment accessory depends on the patient’s age, 

the parameters to be measured and the skin condition at the measurement 

site.  For additional information and detailed sensor application directions please 

refer to the Instruction Manual for the SDMS or the Quick Reference Guide 

of the SDMS or - in case of use with a SenTec-compatible instrument - to the 

instrument’s manual or contact the manufacturer of the instrument.

Terralin Protect

 

(Schülke&Mayr)

22 g Quaternäre Ammo-

niumverbindungen, Ben-

zyl-C12-16-alkyldimethyl-, 

Chloride, 17 g 2-Phenoxy-

ethanol, 0,9 g Aminoalkyl-

glycine, 5 – 15% non-ionic 

tensides.

Carefully wipe all SenTec TC surfaces and the 

cable with wipes soaked in 2% solution (e.g. 980 

ml deionoized water plus 20 ml Terralin Protect). 

Wait 15 min. 

Remove all Terralin residues using 70% Isopropa-

nol. Allow to dry.

Prevantics™ Device 

Swab 

(PDI)

3.15% Chlorhexidine 

Gluconate (w/v) and 70% 

Isopropyl Alcohol (v/v)

Carefully wipe all SenTec TC Sensor surfaces and 

the cable with the pre-soaked PDI wipe for 5 

seconds. Let dry for 5 seconds. Remove all PDI 

residues using 70% Isopropanol. Allow to dry.

Sanoclean AR 

(Sanosil);

1.5% Hydrogen Peroxide 

(0.003% Silver)

Carefully wipe all SenTec TC Sensor surfaces and 

the cable with wipes soaked in Sanoclean AR 

(note: the commercial solution is ready to use, no 

further dilution required), Wait 15 minutes.

Remove all Sanoclean residues using 70% Isopro-

panol. Allow to dry.

Dismozon plus 

(Bode 

Chemie);

Magnesium Monoper-

oxyphthalat 958 mg/g

Carefully wipe all SenTec TC Sensor surfaces and 

the cable with wipes soaked in 3.6% solution (e.g. 

36g of Dismozon granulate per liter of deionized 

water). Note: The diluted solution has a shelf-life 

of 8h. Wait 15 min. 

Remove all Dismozon residues using 70% Isopro-

panol. Allow to dry.

Microbac Forte 

(Bode 

Chemie);

20% Benzyl-C12-

18-alkyldimethylammo-

niumchloride; 5% N-(3-

Aminopropyl)-N-dodecyl-

propan-1,3-diamin

Carefully wipe all SenTec TC surfaces and the 

cable with wipes soaked in 2% solution (e.g. 980 

ml deionoized water plus 20 ml Microbac Forte). 

Wait 15 min. 

Remove all Microbac Forte residues using 70% 

Isopropanol. Allow to dry.

Super Sani Cloth 

(PDI);

0.25% n-alkyl dimethyl 

ethylbenzyl ammonium 

chlorides; 0.25% n-alkyl di-

methyl benzyl ammonium 

chlorides

Carefully wipe all SenTec TC Sensor surfaces and 

the cable with the pre-soaked PDI wipe. Wait 2 

minutes.

Remove all PDI residues using 70% Isopropanol. 

Allow to dry.

      CAUTION:

 This cleang agent is potentially 

aggressive to the skin. Thus, sensor membrane 

must be changed after cleaning to ensure optimal 

skin compatibility.  

CaviCide or CaviWipe

 

(METREX® RESEARCH 

CORPORATION);

0.28% Diisobutyl-

phenoxyethoxyethyl di-

methyl benzyl ammonium 

chloride; Isopropanol

Spray CaviCide or CaviWipe directly onto the 

sensor/cable surface. Allow surface to remain 

visibly wet for 3 minutes. Remove all CaviCide or 

CaviWipe residues using 70% Isopropanol. Allow 

to dry.

      CAUTION:

 This cleang agent is potentially 

aggressive to the skin. Thus, sensor membrane 

must be changed after cleaning to ensure optimal 

skin compatibility.  

Table 2: High Level Disinfection

Brand name/

manufacturer

Cleaning procedure

Cidex OPA

 (J&J);

0.55% 

ortho-Phthalaldehyde

Note: Requires pre-clean-

ing using an enzymatic 

cleaner: 

ENZOL

 (J&J) or 

Prolystica

 (Steris);

Subtilisin

First, remove debris using an enzymatic deter-

gent: Soak the Sensor and the desired length of 

cable in enzymatic solution for 5 minutes. Rinse 

the sensor and the cable briefly in deionized wa-

ter. Blot dry carefully.

Soak the Sensor and the desired length of cable in 

CIDEX OPA solution for 12 minutes.

Soak the SenTec TC Sensor and the desired length 

of cable 3x for 2 minutes in sterile or deionized 

water (renew after each soaking cycle). 

Carefully wipe all surfaces of the SenTec TC Sen-

sor and cable with 70% Isopropanol. Allow to dry.

      CAUTION:

 Sensor membrane must be 

changed after cleaning to ensure optimal sensor 

monitoring performance. 

Reviews: