Sentry Equipment Corp
2.
Fill the container with activated carbon (or impregnated activated carbon) to about 1 in from the top of the
container.
3.
Place approximately 2 in (uncompressed thickness) of glass wool (Sentry part number 4-04825A) at the bottom of
the container.
4.
Replace the container.
5.
Reinstall the clamp.
6.
Place the unit into operation.
Detailed, panel-specific operating instructions are included with this manual.
Troubleshooting
symptom
possible problem(s)
remedy
Elevated sample
temperature
(For cooler option)
Loss of cooling water flow
Increase in cooling water inlet
temperature
Scaled cooler
Increased sample flow
Check cooling water supply
Reduce cooling water temperature or
increase flow
Chemically clean (see cooler maintenance)
Adjust sample flow rate
Reduced sample flow
Plugged line
Blowdown line and exercise flow control
valve
Leaking valve stems
Loose or worn packing
Tighten packing nut or replace packing
Leaking cooler
(For cooler option)
Loose housing from thermal shock
or worn gasket
Tighten bolts on head assembly or replace
gasket if cut or worn
Parts & Accessories
Description
Part Number Quantity
Needle Assembly, .065 in (#16 ga) Needles
2-05513F
1
Needle Assembly, .083 in (#14 ga) Needles
2-05513H
1
Needle Assembly, .110 in (#12 ga) Process and .083 in (#14 ga) Vent Needles 2-05513K
1
Needle Assembly, .083 in (#14 ga) Process and .110 in (#12 ga) Vent Needles 2-05513J
1
Needle Assembly, .065 in (#16 ga) Process and .083 in (#14 ga) Vent Needles 2-05513G
1
Orifice,
.062
in
2-05515A
1
Fluid
Inlet
Port
O-Ring,
Viton
4-04820M
1
Fluid Inlet Port O-Ring, Kalrez
4-04820X
1
Needle
Assembly
O-Ring,
Viton
4-04820L
1
Needle
Assembly
O-Ring,
Kalrez
4-04820Y
1
Sample Bottle, 2 oz (60 ml), Glass with Cap and Septum
4-00775C
1