REFERENCE DATA 4: Range of Liquid Flow Rate
15
Product
Maximum Pressure Resistance
EF-MICRO
0.2 MPa (= 2 bar)
EF-G2 Series
0.2 MPa (= 2 bar)
EF-G3 Series
0.2 MPa (= 2 bar)
EF-G5 Series
0.2 MPa (= 2 bar)
Product
Flow area
EF-MICRO
0.5 ~ 10 ml/min
EF-G2 series
1 ~ 60 ml/min
EF-G3 series
5 ~ 300 ml/min
EF-G5 series
10 ~ 1500 ml/min
15
Make sure that the liquid feeding pressure at the inlet port is lower than the maximum
pressure value described in the specification sheet.
If the feeding pressure exceeds the maximum value, even momentarily, the hollow fiber may
be damaged and ink may leak into the vacuum line.
※
The warranty will not apply in cases where liquid feeding pressure exceeds the maximum
pressure value specified in the specification sheet,
Even if liquid feeding pressure is lower than the specified maximum pressure, an increased
flow rate will cause lesser degassing performance. On the other hand, if the flow rate is too
low, degassing cannot be efficient. Therefore, DIC supplies recommended flow ranges in the
table below.
※
The recommended flow range below was calculated when using water. In the case of
high viscosity liquid, a suitable flow range will be more narrow.
Application Materials Product Division
INSTRUCTION MANUAL