background image

 

Servomex 2500 Quick Start Manual 

2500 Series - Configuration and Operation 02500/003E/0 

29 

NOTE 

  If Auto calibrations are to be remotely initiated, it is advisable to prevent internal initiation 

by setting the auto calibration PERIOD to 00 days, 00 hrs. 

  Timed or remotely-initiated Auto calibrations will NOT operate when there is any Fault 

condition or during the warming-up period, or 15 minutes after the clearance of the last 
Fault or warm-up. 

  An Auto calibration will fail if either the Zero Tolerance or Span Tolerance would be 

exceeded. Any Auto calibration will fail if a Span Preflush is unsuccessful for any 
component. 

  Any Auto calibration in progress can be manually aborted by pressing any key. 

  Some Auto calibration settings may be changed using digital communications where the 

RS485 or Ethernet option has been fitted. See installation manual for details. 

  Any Auto calibration in progress can be manually aborted by pressing any key. It may 

also be aborted by means of a Modbus command. See installation manual for details. 

 

4.5  Calibration when learnt cross interference is enabled 

Where the analyser is fitted with XI (cross interference compensation) and where this is 
enabled, Manual Span calibrations are required with separated span components. 

Auto calibration using a gas mixture is normally unavailable. However if Autocal has 
been ordered then it may be used with a gas mixture using specially determined Span 
Target Values. These values are supplied with the instrument documentation and must 
be used as the Span Target values during span calibration rather than the actual 
mixture composition. 

Summary of Contents for 02500003E

Page 1: ...Part Number 02500003E Revision 0 Language UK English SERVOTOUGH SpectraExact 2500 QUICK START MANUAL...

Page 2: ...Servomex 2500 Quick Start Manual 2 2500 Series Configuration and Operation 02500 003E 0...

Page 3: ...Servomex 2500 Quick Start Manual 2500 Series Configuration and Operation 02500 003E 0 3...

Page 4: ...Servomex 2500 Quick Start Manual 4 2500 Series Configuration and Operation 02500 003E 0...

Page 5: ...nt display screen 16 3 5 Communication Options 17 3 6 Communication Setup 18 3 7 RS232 18 3 8 MODBUS ASCII 20 3 9 MODBUS RTU 20 3 10 ETHERNET 20 3 11 Functions 21 3 12 Associate 23 4 CALIBRATION 24 4...

Page 6: ...the user to pertinent facts and conditions 1 2 About this manual This manual covers software configuration and operation for the 2500 2520 and 2550 range of analysers Further copies of this manual ma...

Page 7: ...nt Display 3 Hard key to select Menu Display 4 Hard key to exit to the previous menu level 5 Fault warning LED 6 Alarm warning LED 7 Hard key to enter data or accept highlighted option within Menu 8 U...

Page 8: ...each level of operation the user is offered a selection or menu of items on the display and he she selects the appropriate one by highlighting it with the cursor keys and then pressing the ENTER key...

Page 9: ...lished While the analyser is warming up the software configuration to suit the user s particular requirements may now be accomplished Proceed to Section 3 Configuration Note that during the warming up...

Page 10: ...arrow means further information in the direction and a double headed arrow indicates more information in either direction 2 3 Setting passwords NOTE If the keyswitch facility is to be used ignore this...

Page 11: ...o retain the time setting for up to 2 days without power At this stage it should be set to the correct local time and date and the preferred day month format in order that all future timed operations...

Page 12: ...o occur iv The HYSTERESIS dead band of the alarm this may be used to avoid chatter if the typical sample concentration is close to the alarm level Go via MENU SET UP to SET ALARM Press ENTER At displa...

Page 13: ...concentration alarms for any of the measured components in any combination ii Calibration in Progress iii Fault Any existing relay allocation may be edited or cleared When cleared the relays will rem...

Page 14: ...ay or select CLEAR to clear the relay then press ENTER 1 3 Measurement AL1 HI 10 0 Y N Select Y to assign this function to this relay or N to clear the function then press ENTER 1 3 Measurement AL2 HI...

Page 15: ...nge Minimum recommended analogue output range is 5 fsd maximum is 200 fsd i e 2x over range Factory default is 0 100 fsd for all outputs NOTE Above 100 fsd linearity performance may be degraded At min...

Page 16: ...1 1 Measurement R1 L 00 0U 20 0 Select Lower and Upper limits for the output and press ENTER 1 1 Measurement R1 TRUE 0 LIVE 0 Select required state then press ENTER 1 1 Measurement R1 FREEZE FOLLOW S...

Page 17: ...s If specified the analyser may be fitted with additional hardware to allow digital communications using either Modbus over RS485 or Modbus TCP over Ethernet If either option is fitted the RS232 outpu...

Page 18: ...TS 8 HANDSHAKE HARDWARE Any device intended for connection to the 2500 must be capable of serial RS232 communication and be configured with these settings Entering the RS232 menu displays further opti...

Page 19: ...Servomex 2500 Quick Start Manual 2500 Series Configuration and Operation 02500 003E 0 19 At display screen Operation SYSTEM SET UP OR IDENTITY OR HISTORY Select required value then press ENTER...

Page 20: ...MODBUS RTU This section details the analyser configuration that is displayed when the communications mode is set to MODBUS RTU This should be adjusted as necessary to match the requirements of the Mod...

Page 21: ...locks out any user adjustments This is useful for applications where span samples are infrequently available e g moisture in solvents iii Zero Constant This should normally be left in USER setting to...

Page 22: ...re fitted ON OFF Select required state then press ENTER SPAN CONST Measurement DEFAULT USER Select required state then press ENTER ZERO CONST Measurement DEFAULT USER Select required state then press...

Page 23: ...essor PCB is being fitted This operation is only carried out once In addition if a Default System Data corruption Fault is diagnosed this may be corrected using the Associate routine Go via MENU SETUP...

Page 24: ...tments are made In either case results are logged in the appropriate Calibration History log Either type of calibration can be performed MANUALLY or AUTOMATICALLY NOTE It is likely that SPAN adjustmen...

Page 25: ...CAL Press ENTER At display screen Operation SELECT COMPONENT CO2vpm if more than 1 component Select required measurement then press ENTER Z CAL S CAL HISTORY CHK Z S Select Z CAL then press ENTER ZER...

Page 26: ...tion has settled then press EXIT the 2500 logs the concentration at this point CHECK ZERO CHECK SPAN Select span if required then press ENTER and repeat the process with the span gas 4 4 Auto calibrat...

Page 27: ...Span sample concentration iv Required Auto calibration period time between cycles v Start time date of the first Auto calibration cycle vi DV dead volume lag This is the time to allow full flushing o...

Page 28: ...ENTER ENTER PERIOD HOURS 0 0 Select required value then press ENTER ENTER START TIME SET YEAR 0000 Select required value then press ENTER ENTER START TIME SET MONTH 00 As above ENTER START TIME SET DA...

Page 29: ...me Auto calibration settings may be changed using digital communications where the RS485 or Ethernet option has been fitted See installation manual for details Any Auto calibration in progress can be...

Page 30: ...o correct any hardware problems arising or deal with any failure of auto calibration sample flow etc ii Serious Faults If any fault arises or parameter degrades to such an extent that the measurement...

Page 31: ...de use the cursor keys to select the next alarm Measurement AL2 ENABLE 10 0 HI repeat above process to view all alarm settings then press EXIT 5 3 Displaying output configuration Relays Go via MENU SE...

Page 32: ...Go via MENU FAULTS to FAULT HISTORY Press ENTER At display screen Operation DET SIG LO ON 17 13 23 23 12 Select the desired fault entry to display then press EXIT to finish NOTE Sometimes one or two...

Page 33: ...asurement CHKMZ 9 2 13 22 25 31 12 Select the desired calibration entry to display then press EXIT to finish 5 5 Displaying analyser identity and diagnostics Identity Go via MENU SET UP DISPLAY to ID...

Reviews: