Installation, Start Up and Shut Down DF-745SG
17
4.2 Analyzer Start Up
It is important to note that, if equipped with automatic gas valves, the moisture cell
is isolated from gas flow while the analyzer is off power.
Open the door and turn on the power using the main power switch inside the
analyzer. See Figure 2. The analyzer will undergo a series of Diagnostic
Procedures while the various startup screens are displayed. Next, the Servomex
Corporation logo is briefly displayed and then the data display appears with the
“Warming Up” screen flashing. The warm up process takes approximately six to
ninety minutes after which the display will look similar to Figure 7 (values shown
are only representative).
During the Warm Up process the gas valves (if equipped) can be opened and gas
flow started. See the section on Gas Pressure and Flow Settings on page 18 for
important information on the gas delivery system and setting proper pressures and
flow rates.
NOTE
During the warm up period all analog and digital outputs
are held to an artificial 0.011ppb reading to avoid the
reporting of false readings.
Figure 7: Data Display Screen
4.2.1 Gas Delivery System
The gas delivery system, as shown in Figure 8, is designed to deliver a gas flow
rate of 1 liter per minute to the moisture cell while maintaining the highest
standards of gas purity and delivery for ultra-trace analysis. Features include a
single inlet line and flow meter for the gas sample, and optional bypass loop to
maintain constant purging, and essentially dead-leg free gas delivery.
Summary of Contents for NanoTrace DF-745SG
Page 1: ...NanoTrace Moisture Analyzer DF 745SG OPERATOR MANUAL ...
Page 4: ......
Page 12: ......
Page 18: ......
Page 28: ......
Page 30: ...24 DF 745SG Options ...
Page 72: ......
Page 74: ......
Page 80: ......
Page 82: ......
Page 86: ...80 DF 745SG Appendix A User Menu Screens Page 47 Page 49 Page 52 Page 53 Page 55 Page 56 ...
Page 88: ...82 DF 745SG Appendix A User Menu Screens ...
Page 93: ......