Connecting to External Devices DF-745SG
25
6 Connecting to External Devices
The analyzer can be interfaced to a variety of external devices via the ports on the
rear panel. Alarm contacts, voltage, and current outputs, and serial
communications are supported. All outputs, analog or digital, are fully isolated
from earth ground.
NOTE
During the six minute warm up period all analog and digital outputs
are held to an artificial 0.011ppb reading to avoid the reporting of
false readings.
J9
J2
J3
J10
J5
J4
SERIAL COMMUNICATION
MOISTURE ANALOG OUTPUT
RELAY OUTPUTS
NOT USED
NOT USED
J8
Figure 10: Rear Panel Electrical Connectors
6.1 Serial Communication Port – J5
Either of two serial communication ports are available at the time of order:
RS232C or RS485 which enable interface between the analyzer and other
operating systems.
Up to 32 units may be accessed via RS-485. Operating parameters are 8 bits, no
parity, and one stop bit. Baud rate may be selected from the menu on the display.
See Table 1 on page 26 for wiring information.