User Interface DF-745SG
37
Figure 20: Isolate Warning
While the moisture cell is isolated from gas flow, a warning will appear at the
bottom of the main display indicating “
Isolated
”.
7.4.3 Restore Sample Gas Flow – Optional
Figure 21: Restore Sample Gas Flow
This command allows the user to return the analyzer gas flow to normal after
isolation.
7.4.4 Calibrate Menu
“Calibration” of the moisture analyzer is somewhat of a misnomer. Because of the
nature of absorption measurements, instrumental drift has minimal effect on the
quantitative moisture content result. Once operating parameters are in place from
the factory, no other
SPAN
“calibration” is necessary.
However, line contributions to offsets in ppb measurements, especially at the sub-
ppb level, are difficult to remove even with a correctly operating system. The
moisture analyzer has an array of zeroing features that enable the user to establish
performance near 0.0 ppb.
Any zero action whether completed or aborted is recorded in the
Zero Cal Log
as
shown in Figure 38.
Use the arrow keys (
and
to scroll up and down through the list. Pressing
ESC
will return to the main display.