40 DF-745SG
User Interface
hit the
Enter
key which will move the highlighted area to the
Apply
button and hit
Enter
to set the value. Use the
Next
key to go back and change the value or move
to the
Done
button, followed by hitting the
Enter
key to leave the screen. Using
the
ESC
key at any time will exit the screen making no changes and return to the
main display.
7.4.4.1.2 Do A Manual Zero
Figure 26: Manual Zero Screen
The manual zero command enables the user to zero the moisture cell in an
interactive manner.
For this purpose, it is necessary to ensure that moisture free gas (sub ppb) is
entering the sensor through the process inlet. When switching to a gas that is
moisture free, it is important to wait a period necessary to allow the reading to re-
stabilize.
It is extremely important to note that moisture is inherently slow to move and as a
result the process of doing a zero action can take many hours before the reading is
at baseline. A premature zero, although acceptable, will ultimately result in a
negative reading as the plumbing and analyzer continue to dry out requiring
another zero action. See the section on Active Zero on page 55 for additional
information.
After selecting
Do a Manual Zero
, a screen will appear which displays a trace of
the recent moisture reading. See Figure 26. Observe the trace until the reading is
stable and then press
Enter
. This action will accept the present value as the new
zero setting for the moisture cell and the
Zero Reference
field will be updated.
After this action, the user will be brought back to the Check/Adjust Zero screen.
Pressing ESC during the calibration process will abort the action and return to the
main display. Whether the calibration is complete or aborted, pressing
Done
from
the previous screen will allow the user to return to the main display.
Once complete the gas source must be returned to the original state with process
gas going directly through the sensor.
Summary of Contents for NanoTrace DF-745SG
Page 1: ...NanoTrace Moisture Analyzer DF 745SG OPERATOR MANUAL ...
Page 4: ......
Page 12: ......
Page 18: ......
Page 28: ......
Page 30: ...24 DF 745SG Options ...
Page 72: ......
Page 74: ......
Page 80: ......
Page 82: ......
Page 86: ...80 DF 745SG Appendix A User Menu Screens Page 47 Page 49 Page 52 Page 53 Page 55 Page 56 ...
Page 88: ...82 DF 745SG Appendix A User Menu Screens ...
Page 93: ......