46 DF-745SG
User Interface
The Moisture Zero Log reports on adjustments made to the moisture cell zero
setting. The date and time of the zero calibration is noted. The zero is listed as
either Manual or Automatic. Notes are also given as to if the zero was aborted,
failed due to timeout, or in the event of an Automatic Zero, if it were scheduled.
The time to perform the zero and the resulting zero reference are also noted.
Figure 38: Zero Log Screen
7.4.7.2 System Error Code Log
Figure 39: View System Error Code Log Menu
The System Error Log reports functional errors in the moisture system. If the error
persists for more than 30 minutes, the code is displayed, if warranted. In addition,
a system alarm will trip if configured to do so. See page 51 for additional
information on setting System Alarms.
NOTE: The 30 minute clock is delayed for 60 minutes on a “cold” system start up.
Following is a list of System Error/Event Codes and their descriptions:
101 = peak unstable or not found
133 = data acquisition system event