50 DF-745SG
User Interface
repaired. A warning describing this condition will appear over the main display and
the user will be unable to restore gas flow until the fan is repaired. See Fan Failure
on page 59.
7.4.8.2 Moisture Alarm Setup
The Setpoint value refers to the limit above or below which the alarm is triggered.
The Trip command sets the above/below parameter. The deadband refers to the
value from the nominal setpoint that the output value must exceed before an alarm
is reset. The Relay Assignment indicates to which relay the alarm is assigned.
Use the
Next
key to move from between fields and use the arrow keys (
and
to change the highlighted selections and to enter numerical values. When
done, use the
Next
key to move to the
Accept
button and hit the
Enter
key to
return to the main display. Using the
ESC
key at anytime will exit the screen
making no changes and return to the main display.
Figure 45: Alarm Setup Screen
7.4.8.3 Temperature Range Alarm Setup
Figure 46: Temperature Alarm Setup