User Interface DF-745SG
53
range is required, rather than two, then the value of FS B should be set to equal
the FS A value.
Figure 50: Analog Output Setup Screen
A window as narrow as 10% of the analyzer’s decades can be set for the full-scale
analog output.
The
In Calibration Relay
can be enabled to signal that a zero calibration is in
process. In addition, the user has the option to freeze the analog output or enable
the analog output to update as the calibration progresses.
7.4.10
Graph Setup
Figure 51: Graph Setup Menu
The graph setup is used to adjust the time scale on the main data display of the
analyzer. A specific time interval in minutes can be chosen for the X-axis on that
display. The minimum acceptable time is 3 minutes. The information on the
display represents current data and will show a history of moisture concentration
based on the given time span to the present. The Y-axis of the main data display is
auto-ranging.