32
5 Testing / maintenance
5.1
Function testing, testing display accuracy, maintenance
Please consider the necessary and prescribed instruments con-
trol according to:
EN 50073
EN 45544-4
BGI 518 (T023)
BGI 836 (T021)/Applicable only in Germany
DVGW worksheet G 465-4/Applicable only in Germany
Testing must also cover accessories used. Tests carried out and
other activities must be documented and the documentation re-
tained for at least one year. The instrument tests required and
prescribed by DVGW G 465-4 (Technical Communications, re-
ference) are divided into the following sections:
What ?
Who ?
When ?
function testing
user
prior to start working
testing display accu-
racy (adjustment)
specialist or special-
ist company
daily to half-yearly
servicing
(maintenance, any
repairs required)
SEWERIN,
specialist or author-
ised company
annual;
on defective instru-
ments
Function testing
This is the simplest form of instrument test, which must be car-
ried out by the
user
before starting work. It covers the following
items:
external condition including probe systems
function-testing the controls
battery charge
inspecting the pump and suction inlet
pump function
zero point
display of test gas feeding
Summary of Contents for EX-TEC Combi
Page 1: ...Operating Instructions ...
Page 3: ...Illustration EX TEC Combi EX TEC Combi Device variation 1 Device variation 2 EX TEC Combi 100 ...
Page 5: ...Operating Instructions 15 06 2005 V2 X 103401 en EX TEC Combi ...
Page 117: ...108 Appendix ...
Page 118: ...109 Appendix ...
Page 119: ...110 Appendix ...
Page 120: ...111 Appendix ...
Page 121: ...112 Appendix ...
Page 122: ...113 Appendix ...
Page 123: ...114 Appendix ...
Page 124: ...115 Appendix ...
Page 125: ...116 Appendix ...
Page 126: ...117 Appendix ...