39
5 SePem 01 Master
Data / Results
Data/Results
in the menu provides information from the logger.
The logger data specified in
(first table) is listed.
Please note that some values are only displayed if the logger has
saved corresponding measurements.
Any
logger errors
detected during read-out will be shown under
General
in the last line. The error number given will be important
for any contact with SEWERIN Service.
Online measurement
When online measurement is active the measurement data re-
corded by the logger can be viewed directly on the
Master
in
real time.
Note:
Online measurements cannot be saved.
Online measurement is typically used during
mobile monitoring
.
This is useful for providing feedback on the noise levels detected
by the logger immediately after installation. If the level is close to
zero then it will not be necessary to monitor the installation location
with a stationary unit. The logger can be removed immediately
and reinstalled at another location.
Usually, measurement values are calculated every second. All
measurement values are immediately displayed in a chart. The
F2
Stop
key can be used to stop the measurement in order to view
the graphics. (Note: The measurement
cannot
be subsequently
restarted.) Pressing
F3
Back
cancels the online measurement.
Please note the following points:
●
The measurement interval (
Logger configuration
>
Meas.
times
>
Interval
) specifies the length of time between each
measurement that is recorded, and therefore determines the
speed at which the graphics is updated.
●
If an online measurement is running at the same time as a
predefined measurement then the scale of the time axis will
depend on the measurement interval.