Page 34 of 40
METER -
advanced confi guration
9.6.2 - MEASURE STATUS
Position the cursor on MEASURE STATUS, press ENTER to access.
It’s possible to display the gain of the system, with values from 0 to 255.
LEFT ARROW to exit
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
►
G: 00000
MEASURE STATUS
9.6.3 - FROZEN GAIN
Position the cursor on MEASURE STATUS, press ENTER to access.
It’s possible to fi x a value of gain (from 1 to 255) and consequently disa-
ble the automatic gain control. Once the value is 000 the automatic gain
control restarts.
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: 000
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
►
000
FROZEN GAIN
9.6.4 - MAX GAIN TH
Position the cursor on MAX GAIN TH, press ENTER to access.
It’s possible to change the max value of gain. If the gain reaches this value,
the “GAIN” error code is activated
Use UP ARROW and SCROLL to modify the value.
Press ENTER to confi rm.
LEFT ARROW to exit without changes
Default value: 200
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
►
200
MAX GAIN TH
9.6.5 - PEAK VALUES
Position the cursor on PEAK VALUES, press ENTER to access.
The system store the maximum distance and the minimum distance meas-
ured since the power is turned ON.
It’s possible to see those values or reset the values.
With the SCROLL button you can select the function.
Press ENTER to confi rm.
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
►
DISPLAY VALUES
RESET VALUES
►