1ADJUSTMENTS, PERFORMANCE CHECK, AND FIRMWARE UPGRADE
2 – 4
CONFIDENTIAL
3.
The communicating dialog box appears while processing.
4.
When the restore is complete, click "OK".
4-4. DOWNLOADER
4-4-1. Introduction
Downloader allows you to upgrade the firmware.
1)
System requirements
• Requirements for the upgrading tool:
• Requirements for Loading loader
2) Required devices
The following devices are required to rewrite MOT files using Communi-
cation Box.
Conversion connector (16-pin
3
10-pin)
This conversion connector is required to connect TM200 with Communica-
tion Box since the cable connector of Communication Box has 16 pins.
4-4-2. Getting started
This section describes how to install/uninstall the software and how to
connect handset to a PC.
1)
Installing/uninstalling the software
[Installing the software]
Note: Downloader is installing/uninnstalling simultaneously with SPTP.
1. Double click the “setup.exe” icon in TM200 Install-E folder.
2. The Setup Wizard is activated.
Click the "Next" button.
3. A location to install the software appears.
To install in the default location, click the "Next" button.
To change the location, click the "Browse" button, select a desired
location and then click the "Next" button.
(See the description in "* Changing the install location" below.)
OS
: Windows 98, 98 SE, 2000, or XP
Download file
: Only the Motorola format is supported.
Communication method : Asynchronous mode
Data length
: 8 bits
Parity bit
: None
Stop bit length
: 1 bit
Handset
: TM200
Communication method
: Asynchronous mode
Data length
: 8 bits
Parity bit
: None
Stop bit length
: 1 bit