6
4. Cautions and Notices
Observe the
following:
CAUTION!
Use only the AC adaptor provided by the distributor authorized
by Shimadzu if using an AC adaptor.
NOTICE
Make sure that any foreign objects, including water and metal
pieces, do not enter the interior of the balance.
NOTICE
Do not leave the balance with objects on the pan.
NOTICE
Do not cause a shock to the pan.
5. Measurement Procedure
The procedures are explained using the ELB300 as an example.
Turn on the power
(1)
(2)
(3)
(4)
(5)
(6)
Check that nothing is placed on the pan.
Press the [ON/OFF(BREAK)] key.
The Check display will illuminate, indicating
that the temperature sensor inside the
balance is being checked.
Depending on the ambient temperature, this
may remain illuminating for about 30 seconds.
The all segments illuminate for approximately
6 seconds.
Zero is displayed.
The zero may be unstable immediately after
the power is turned on. Allow the balance to
warm up for more than 5 minutes. To conduct
accurate measurements, a warm up time of
over 30 minutes is recommended.
Calibrate sensitivity (refer to “9. Calibration”).