Place the container* on the center of the pan
and press the [TARE] key.
Check that the Center Zero mark illuminates.
Load the sample and read the displayed
value when the display stabilizes.
To conduct accurate measurements, perform
a preliminary loading* before conducting
measurements.
“
oL
” will be displayed if the weight of the
sample exceeds the capacity, or the
combined weight of the sample and
container exceeds the total available range
of the balance.
Dry Batteries consumption can be reduced by activating the Auto Power Off
function.
Refer to “6. Auto Power Off Function” for information on how to set and disable the
Auto Power Off function.
*Container:
Any container in which the sample is placed. Measurements without any container
are also possible.
*Preliminary
loading:
“Preliminary loading” refers to loading and unloading the sample or the weight
once before conducting the actual measurement. A preliminary loading improves
the accuracy of the measurement.
Turn off the power
Press the [ON/OFF(BREAK)] key during weight display.
CAUTION!
Be sure to turn off the power when the balance is not to be
used for a long period of time.