If the following codes display during operation, solve the problem by applying the solution below.
Code Description
Solution
C E4
This appears when an
improper weight is placed on
the pan during calibration.
Restart calibration with the correct weight.
C E2
This appears when trying to
start calibration with an object
on the pan.
Restart calibration after removing the object from the
pan.
r E4
This appears when the
reference value is smaller than
the minimum increment in
piece counting or percentage
conversion mode.
Reset the reference with a larger amount of sample.
Err 1
Internal temperature sensor
error.
Err 4
Internal computation error.
Stop using the balance and contact our service
personnel.
17. Specifications
Model
ELB120
ELB200 ELB300 ELB600 ELB1200
Weighing
capacity
120g 200g 300g 600g 1200g
Minimum display
0.01g 0.01g 0.01g 0.05g 0.1g
Standard deviation
σ≤
0.01g
σ≤
0.01g
σ≤
0.01g
σ≤
0.05g
σ≤
0.1g
Linearity errors
±
0.01g
±
0.01g
±
0.02g
±
0.05g
±
0.1g
Calibration
weight 100g 200g 300g 500g 1000g
Pan size (mm)
φ
110
170x130
Stability of Sensitivity
( 5 to 35
°
C )
±
20ppm/
°
C
±
15ppm/
°
C
±
10ppm/
°
C
±
20ppm/
°
C
±
20ppm/
°
C
Dimensions of main
body (mm)
Approx. 185(W) x 215(D) x 55(H)
Weight (kg)
Approx. 1.25
Temperature range
5 to 40
°
C
Power supply
6 x AA dry batteries
9DCV (1.5DCV x 6)
Continuous
operation time
Approx. 40 hours
(using alkaline dry batteries)
AC adapter
(Option)
DC output voltage: 9 DCV
DC output current: 80mA
Shape of the plug: