28
19.2.3 Input/Output Format
Key: “
∆
” indicates a space and (CR) indicates a carriage return.
Input data
Command code + (CR) -> See “19.2.4 Command Codes”.
Output data
- When displaying the weight
S -
∆∆
300.00 g
∆
(CR)
Unit + space
Space, number, decimal point
Polarity Positive
space
(
∆
)
Negative
minus (-)
Internal stability
(when outputting with stability information)
Stable
S
Unstable
U
- When “
oL
” or “
-OL
” is displayed
S -
∆∆∆∆
oL
∆∆∆∆
(CR)
Unit + space
Polarity Positive
space
(
∆
)
Negative
minus
(-)
Internal stability
(when outputting with stability information)
Stable
S
Unstable
U
Data format
- ASCII (JIS) code
- Baud rate
Select from 300, 600, 1200, 2400, 4800, and 9600.
- Parity
Select from “
EvEn
” (even number), “
odd
” (odd number), and
“
non
” (none).
- Data length
Without parity
8 bit
With parity
7 bit
- Stop bit
Select from 1 and 2.
NOTICE
Set the baud rate to 1200 and the stop bit to 1 when using the EP-80.