4.2.4 Test Mode

Change the mode of operation between the four modes. 
Press “ZERO” or “MODE” keys to select. Press “LOG” to can-
cel or “MENU” to confirm and exit. This adjustment can also 
be changed at the home screen display by simply pressing 
“MODE” (Figure 4-2f).

 

First Peak Mode will display a drop ra-

tio menu (Figure 4-2g). This drop ratio actives the first peak 
recording.

          Figure 4-2f                                   Figure 4-2g
4.2.5 Peak Time

In the Peak Time menu, the peak auto reset time can be set. 
The range is 1-99 seconds. Touch “ZERO” to adjust the value, 
press “MODE” to shift to the next position. Press “LOG” to can-
cel; touch “MENU” to confirm and exit. (Figure 4-2h)

Figure 

4-2h

4.2.6 Alarm

The alarm function can be turned on or off to activate or de-
activate the user programmed tolerances set in the Tolerance 
Menu. Touch “ZERO” or “MODE” keys to shift to the next po-
sition. Press “LOG” to cancel, touch “MENU” to confirm and 
exit. (Figure 4-2i)

Figure 4-2i

4.3 Memory 

In the Memory menu, the user can select the mode of data stor-
age, browse, delete, or print the data. (Figure 4-3a)

Figure 4-3a 

 

4.3.1 Storage Mode

Two storage modes can be selected in this menu: Single and 
Series. Touch “ZERO” or  “MODE” keys to select between the 
two. Press “LOG” to cancel; touch “MENU” to confirm and exit. 
(Figure 4-3b)

Single:

 At the home screen, pressing the “LOG” stores the cur-

rent displayed value. (If the default settings key is for storage. 
See 4.5.8 key setting.)

Series:

 Continuous data logging will only operate while in the 

Auto Peak measuring mode. When the peak time has expired, 
unit stores the current displayed peak value and then resets 
the peak value on the display. Touch “LOG” to start, touch
“LOG” again to end. 

Figure 4-3b

4.3.2 Browse All

The data will be displayed. Touch “ZERO” or “MODE” keys to 
shift to the next position. Touch “MENU” to see Delete
or Print options. Touch “LOG” to go back. (Fig. 4-3c)

1

 Position number  

2

 Data and units      

3

 Force Direction      

4

 First Position Data

Figure 4-3c

4.3.4 Browse Selected

User can choose the data to browse. The available range of 
data stored is shown. Touch “ZERO” to adjust the value. Press     
“MODE” to shift to the next position. Press “LOG” to cancel; 
touch “MENU” to confirm. (Figure 4-3d)

Figure 4-3d

4.3.5 Delete Selected

Select the range of data to be deleted. Touch “ZERO” to adjust 
the value. Press “MODE” to shift to the next position. Touch       
“LOG” to cancel; touch “MENU” to confirm. (Figure 4-3e)

Figure 4-3e

4.3.6 Delete All

In this menu, a prompt will appear.  All data will be deleted by 
selecting “YES” and canceled by selecting “NO” or pressing 
“LOG”. (Figure 4-3f)

Figure 4-3f

4

1

2

3

4

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for FG-7000L Series

Page 1: ...O INSTRUMENTS Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test Whether the unit is ON or OFF DO NOT exceed the capacity of the sensor NEVER exceed 150 of the rated capacity for S beam or 120 for ring type or the load cell will be damaged At 110 of the rated capacity the display will flash a warning When mounting FG 7000L Se...

Page 2: ... 10 Resolution 1 0 1 0 5 0 01 S 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 Ring Type for Compression N kgf lbf kN klbf tf R 10 10kN Capacity 10000 1000 2250 10 Resolution 0 1 0 1 0 5 0 01 R 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 R 50 50kN Capacity 5000 50 11 5 Resolution 1 0 01 0 001 0 001 R 100 100kN Capacity 100 22 10 Resolution 0 01 0 005 0 001 R 200 200kN Capacity 200 4...

Page 3: ... selec tion Press LOG to cancel or touch MENU to confirm and exit Figure 4 2c Figure 4 2c 4 2 2 Group When several test samples need to be measured the sam ples can be coded into groups The range is 01 99 When set to 00 become 01 automatically Press ZERO to adjust the value touch MODE to shift to the next position Touch LOG to cancel press MENU to confirm and exit Figure 4 2d Figure 4 2d 4 2 3 Tol...

Page 4: ...LOG to cancel touch MENU to confirm and exit Figure 4 3b Single At the home screen pressing the LOG stores the cur rent displayed value If the default settings key is for storage See 4 5 8 key setting Series Continuous data logging will only operate while in the Auto Peak measuring mode When the peak time has expired unit stores the current displayed peak value and then resets the peak value on th...

Page 5: ...menu several parameters may be set per Fig ure 4 5a 4 5b Figure 4 5a Figure 4 5b 4 5 1 Display Mode Two display modes may be selected Digital and Graphic Fig ure 4 5c Figure 4 5c 4 5 2 Display Direction Select the mode of the LCD display Automatic Obverse and Reverse Touch ZERO or MODE keys to shift to the next position Press LOG to cancel Push MENU to confirm and exit Figure 4 5d Figure 4 5d 4 5 ...

Page 6: ...perate improperly you can restore it to the factory default settings Carefully use this function Figure 4 5l Figure 4 5l 4 6 Language Select between English German and Chinese Figure 4 6a Figure 4 6a 4 7 Calibration Users can field calibrate the gauge First enter the system pass word Default is 123 by pressing the ZERO and MODE keys Then press MENU to confirm Figure 4 7a Figure 4 7a 1Calibration P...

Page 7: ...dard protocol Figure 6 1a The USB Port can be connected to a charger with the USB cable for charging the internal Ni MH battery and can be connected to a computer for uploading the measured values Connect the gauge and the computer with the USB cable then open the computer software Upload the values Please refer to the software manual for addi tional information 6 2 Port Pin Assignments Figure 6 1...

Page 8: ...Communications Port See Figure 6 1a 7 3 Dimensions S Beam Ring Type Capacity Dimensions W H B M 1kN 51 76 2 19 M12X1 75 2kN 5kN 10kN 25 4 20kN 76 2 108 M18X1 5 Capacity Dimensions mm D d L H SR DI M 10kN 32 8 14 16 16 24 5 3 M5 20kN 38 11 30 50kN 82 22 32 44 50 68 4 M8 100kN 200kN 100 28 36 48 120 80 S Beam Load Cell Ring Load Cell www com information itm com 1 800 561 8187 ...

Reviews: