     4.5.6 Date/Time

The system time may be set under this menu. Touch “ZERO”       
to adjust the value. Press “MODE” to shift to the next position. 
Touch “LOG” to cancel. Press “MENU” to confirm and exit. 
(Figure 4-5h)

    

Figure 4-5h

     4.5.7 Password

The system password can be changed. First, enter the old 
password, then enter the new password and confirm the 
new password. The default System password is “123”. Touch 
“ZERO” to adjust the value. Press “MODE” to shift to the next 
position. Touch “LOG” to cancel; Push “MENU” to confirm and 
exit. (Figure 4-5i)

Figure 4-5i

4.5.8 Key Setting

Set the default function of the “LOG” key from the home 
screen.  The function can be set to print or store the current 
displayed value. Press “ZERO” or “MODE” to select the proper 
setting.  Press “LOG” to cancel; touch “MENU” to confirm and 
exit. (Figure 4-5j)

 

 

 

Figure 4-5j

4.5.9 RS232 Baurate

Adjust Baurate to available bits per second selection in Figure 
4-5k.

Figure 4-5k

4.5.10 Default Setting

If you make a selection that you feel has caused the gauge 
to operate improperly, you can restore it to the factory default 
settings. Carefully use this function! (Figure 4-5l)

 Figure 4-5l

 

4.6 Language

Select between English, German and Chinese (Figure 4-6a)  

        Figure 4-6a

4.7 Calibration

Users can field-calibrate the gauge. First, enter the system pass-
word (Default is 123) by pressing the “ZERO” and  “MODE” keys. 
Then press “MENU” to confirm. (Figure 4-7a)

Figure 4-7a

1

Calibration Point  

2

Current Value      

3

Standard Input Value

      Figure 4-7b

Load a standard force on the gauge. Wait a moment for the force 
to stabilize. The current value (2) should equal the standard force 
applied. 

If the values do not match, press “ZERO” and “MODE” keys to 
correct the standard input value (3).

Press “MENU” to enter the next calibration point. After any of the 
calibration points have been completed, touch “LOG” to exit the 
calibration mode. Then save the calibration or discard by pressing 
“Yes” or “No”.

After completing the calibration of the 5th point, the confirmation 
window will automatically ask to “Save and Exit” by selecting 
“Yes” or “No”. (Figure 4-7c)

Figure 4-7c

Press “ZERO” or “MODE” to select, then press “MENU”. If “Yes” 
is selected, “Calibration Complete!” is displayed.

6

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for FG-7000L Series

Page 1: ...O INSTRUMENTS Operators should wear protection such as a mask and gloves in case pieces or components break away from the unit under test Whether the unit is ON or OFF DO NOT exceed the capacity of the sensor NEVER exceed 150 of the rated capacity for S beam or 120 for ring type or the load cell will be damaged At 110 of the rated capacity the display will flash a warning When mounting FG 7000L Se...

Page 2: ... 10 Resolution 1 0 1 0 5 0 01 S 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 Ring Type for Compression N kgf lbf kN klbf tf R 10 10kN Capacity 10000 1000 2250 10 Resolution 0 1 0 1 0 5 0 01 R 20 20kN Capacity 2000 4500 20 Resolution 0 5 1 0 005 R 50 50kN Capacity 5000 50 11 5 Resolution 1 0 01 0 001 0 001 R 100 100kN Capacity 100 22 10 Resolution 0 01 0 005 0 001 R 200 200kN Capacity 200 4...

Page 3: ... selec tion Press LOG to cancel or touch MENU to confirm and exit Figure 4 2c Figure 4 2c 4 2 2 Group When several test samples need to be measured the sam ples can be coded into groups The range is 01 99 When set to 00 become 01 automatically Press ZERO to adjust the value touch MODE to shift to the next position Touch LOG to cancel press MENU to confirm and exit Figure 4 2d Figure 4 2d 4 2 3 Tol...

Page 4: ...LOG to cancel touch MENU to confirm and exit Figure 4 3b Single At the home screen pressing the LOG stores the cur rent displayed value If the default settings key is for storage See 4 5 8 key setting Series Continuous data logging will only operate while in the Auto Peak measuring mode When the peak time has expired unit stores the current displayed peak value and then resets the peak value on th...

Page 5: ...menu several parameters may be set per Fig ure 4 5a 4 5b Figure 4 5a Figure 4 5b 4 5 1 Display Mode Two display modes may be selected Digital and Graphic Fig ure 4 5c Figure 4 5c 4 5 2 Display Direction Select the mode of the LCD display Automatic Obverse and Reverse Touch ZERO or MODE keys to shift to the next position Press LOG to cancel Push MENU to confirm and exit Figure 4 5d Figure 4 5d 4 5 ...

Page 6: ...perate improperly you can restore it to the factory default settings Carefully use this function Figure 4 5l Figure 4 5l 4 6 Language Select between English German and Chinese Figure 4 6a Figure 4 6a 4 7 Calibration Users can field calibrate the gauge First enter the system pass word Default is 123 by pressing the ZERO and MODE keys Then press MENU to confirm Figure 4 7a Figure 4 7a 1Calibration P...

Page 7: ...dard protocol Figure 6 1a The USB Port can be connected to a charger with the USB cable for charging the internal Ni MH battery and can be connected to a computer for uploading the measured values Connect the gauge and the computer with the USB cable then open the computer software Upload the values Please refer to the software manual for addi tional information 6 2 Port Pin Assignments Figure 6 1...

Page 8: ...Communications Port See Figure 6 1a 7 3 Dimensions S Beam Ring Type Capacity Dimensions W H B M 1kN 51 76 2 19 M12X1 75 2kN 5kN 10kN 25 4 20kN 76 2 108 M18X1 5 Capacity Dimensions mm D d L H SR DI M 10kN 32 8 14 16 16 24 5 3 M5 20kN 38 11 30 50kN 82 22 32 44 50 68 4 M8 100kN 200kN 100 28 36 48 120 80 S Beam Load Cell Ring Load Cell www com information itm com 1 800 561 8187 ...

Reviews: